Abstract

In magnetic resonance electrical impedance tomography (MREIT) we inject currents through electrodes placed on the surface of a subject and try to reconstruct cross-sectional resistivity (or conductivity) images using internal magnetic flux density as well as boundary voltage measurements. In this paper we present a static resistivity image of a cubic saline phantom (50 x 50 x 50 mm3) containing a cylindrical sausage object with an average resistivity value of 123.7 ohms cm. Our current MREIT system is based on an experimental 0.3 T MRI scanner and a current injection apparatus. We captured MR phase images of the phantom while injecting currents of 28 mA through two pairs of surface electrodes. We computed current density images from magnetic flux density images that are proportional to the MR phase images. From the current density images and boundary voltage data we reconstructed a cross-sectional resistivity image within a central region of 38.5 x 38.5 mm2 at the middle of the phantom using the J-substitution algorithm. The spatial resolution of the reconstructed image was 64 x 64 and the reconstructed average resistivity of the sausage was 117.7 ohms cm. Even though the error in the reconstructed average resistivity value was small, the relative L2-error of the reconstructed image was 25.5% due to the noise in measured MR phase images. We expect improvements in the accuracy by utilizing an MRI scanner with higher SNR and increasing the size of voxels scarifying the spatial resolution.

摘要

在磁共振电阻抗断层成像(MREIT)中,我们通过放置在受试者体表的电极注入电流,并利用内部磁通密度以及边界电压测量来重建横截面电阻率(或电导率)图像。在本文中,我们展示了一个边长为50×50×50立方毫米的立方体盐水质模的静态电阻率图像,该质模包含一个平均电阻率值为123.7欧姆·厘米的圆柱形香肠状物体。我们当前的MREIT系统基于一台实验性的0.3T磁共振成像扫描仪和一个电流注入装置。我们在通过两对表面电极注入28毫安电流的同时,采集了质模的磁共振相位图像。我们从与磁共振相位图像成比例的磁通密度图像中计算出电流密度图像。利用J替代算法,根据电流密度图像和边界电压数据,我们在质模中部38.5×38.5平方毫米的中心区域内重建了一幅横截面电阻率图像。重建图像的空间分辨率为64×64,香肠的重建平均电阻率为117.7欧姆·厘米。尽管重建平均电阻率值的误差很小,但由于测量的磁共振相位图像中的噪声,重建图像的相对L2误差为25.5%。我们期望通过使用具有更高信噪比的磁共振成像扫描仪并增加牺牲空间分辨率的体素大小来提高精度。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验