Panisset F, Treffel P, Faivre B, Lecomte P B, Agache P
Service de Dematologie I, Hôpital St Jacques, Besançon, France.
Acta Derm Venereol. 1992;72(1):4-5.
The aim of this study was to demonstrate differences in human skin transepidermal water loss (TEWL) in vivo related to site on volar forearm. Fourteen healthy volunteers entered the study and seven sites were tested. After randomization of both forearm and measure order, TEWL measurements were performed using the Servo Med Evaporimeter. TEWL values next to the wrist were found statistically greater than on the other sites. Hence wrist region should preferably be excluded from TEWL measurements on forearm.
本研究的目的是证明人体皮肤经表皮水分流失(TEWL)在体内与掌侧前臂部位的差异。14名健康志愿者参与了该研究,并对7个部位进行了测试。在对前臂和测量顺序进行随机化后,使用Servo Med蒸发仪进行TEWL测量。发现手腕旁边的TEWL值在统计学上高于其他部位。因此,在前臂进行TEWL测量时,最好排除手腕区域。