Brunmark P, Persson P, Skarping G
Department of Occupational and Environmental Medicine, University Hospital, Lund Sweden.
J Chromatogr. 1992 Sep 2;579(2):350-4. doi: 10.1016/0378-4347(92)80403-d.
4,4'-Methylenedianiline was determined in human urine by micro liquid chromatography with ultraviolet detection. The combination of a thorough work-up and the high mass sensitivity of micro liquid chromatography gave the method very high sensitivity. Derivatization with pentafluoropropionic anhydride enhanced the resolution of the 4,4'-dimethylenedianiline peak. The detection limit, defined as blank plus three times the standard deviation of the blank, was 2 nmol/l of urine, for 10-microliters injection volumes. The detection limit, defined as three times the noise, was about ten times better. The within- and between-assay coefficients of variation were 4 and 6%, respectively, for samples containing 40 nmol/l. The method was applied for the monitoring of excreted 4,4'-methylenedianiline in urine, during epicutaneous skin hypersensitivity testing (patch testing).