Modification of a commercial haemagglutination assay for the detection of anti-HBs has reduced the test time from over one hour to approximately 25 min. increased sensitivity ten-fold without any prozoning, maintained specificity and reduced costs by 90%. The modification consists of diluting the reagent cells ten-fold; these are then added to dilutions of test serum in a V-well microplate. After incubation, plates are centrifuged and then inclined at 70 degrees. Positive and negative reactions can be clearly distinguished within approximately 10 minutes.