Contact versus immersion biometry of axial length before cataract surgery.

作者信息

Hennessy Michael P, Chan Derek G

机构信息

Department of Ophthalmology, Prince of Wales Hospital, NSW 2031 Randwick, Australia.

出版信息

J Cataract Refract Surg. 2003 Nov;29(11):2195-8. doi: 10.1016/s0886-3350(03)00224-4.

Abstract

PURPOSE

To compare the repeatability and agreement of contact and immersion ultrasound (US) biometry of axial length.

SETTING

Prince of Wales Hospital, Sydney, Australia.

METHODS

This prospective stratified randomized study comprised 36 patients with a preoperative refractive error less than 4.00 diopters (D) sphere or 2.00 D cylinder who were recruited before cataract surgery. Each of 3 operators measured both eyes of 12 patients by contact and immersion US techniques. A repeat measurement by both techniques was performed by the same operator in 6 cases and by a different operator in the other 6. Repeat and operator effects were examined.

RESULTS

The mean and standard deviation of the measurement sets were compared, and the differences between repeat measures were calculated. Axial length measurement was longer with the contact method than with immersion by 0.03 mm (P =.04). The repeatability of the 2 techniques was similar.

CONCLUSIONS

When the measurement set was repeated, the precision of contact US biometry was comparable to that of immersion, with no clinically significant difference in mean axial length measurements.

摘要

目的

比较接触式和浸没式超声测量眼轴长度的重复性和一致性。

设置

澳大利亚悉尼威尔士亲王医院。

方法

这项前瞻性分层随机研究纳入了36例白内障手术前屈光不正小于4.00屈光度(D)球镜或2.00 D柱镜的患者。3名操作人员分别采用接触式和浸没式超声技术测量12例患者的双眼。其中6例由同一名操作人员采用两种技术进行重复测量,另外6例由不同操作人员进行重复测量。分析重复测量和操作人员的影响。

结果

比较测量组的均值和标准差,并计算重复测量之间的差异。接触式测量的眼轴长度比浸没式长0.03 mm(P = 0.04)。两种技术的重复性相似。

结论

重复测量时,接触式超声生物测量的精度与浸没式相当,平均眼轴长度测量无临床显著差异。