Gilbert D A, DiMarino A J, Jensen D M, Katon R, Kimmey M B, Laine L A, MacFadyen B V, Michaletz-Onody P A, Zuckerman G
Gastrointest Endosc. 1992 Nov-Dec;38(6):753-6.
Monopolar hot biopsy forceps were developed for simultaneous tissue biopsy and electrocoagulation. Many endoscopists used these forceps for coagulation of diminutive polyps of the colon. The rationale for diminutive polyp eradication is to destroy neoplastic tissue and possibly prevent colon cancer. However, convincing data to document a reduction in the incidence of colorectal cancer or even complete obliteration of all treated diminutive polyps with hot biopsy forceps are lacking. Complications of hot biopsy include hemorrhage, perforation, and post-coagulation syndrome. Tissue injury is deeper with monopolar hot biopsy forceps than bipolar forceps. The right colon is particularly susceptible to transmural injury and perforation. For small polyp obliteration, comparative studies of hot biopsy (monopolar and bipolar) with other techniques such as cold biopsy combined with thermal probes, large cup cold biopsy removal, and snare electrocoagulation are warranted. The necessity to biopsy typical appearing angiomata does not seem warranted on a routine clinical basis. The expected obliteration rates of small angiomata or rates of controlling lower gastrointestinal bleeding from colon angiomata after monopolar hot biopsy electrocoagulation have not been well documented. Heater probe or bipolar electrocoagulation have been safely and effectively applied to bleeding colon angiomata. These newer coagulation probes are recommended as an alternative to hot biopsy forceps for treatment of bleeding colonic angiomata.
单极热活检钳是为同时进行组织活检和电凝而研发的。许多内镜医师使用这些钳子对结肠微小息肉进行电凝。根除微小息肉的基本原理是破坏肿瘤组织并可能预防结肠癌。然而,缺乏令人信服的数据来证明热活检钳能降低结直肠癌的发病率,甚至完全消除所有经治疗的微小息肉。热活检的并发症包括出血、穿孔和电凝后综合征。单极热活检钳造成的组织损伤比双极钳更深。右半结肠尤其容易发生透壁损伤和穿孔。对于小息肉的消除,有必要对热活检(单极和双极)与其他技术,如冷活检联合热探头、大杯冷活检切除和圈套电凝进行比较研究。在常规临床基础上,对典型外观的血管瘤进行活检似乎没有必要。单极热活检电凝后,小血管瘤的预期消除率或控制结肠血管瘤引起的下消化道出血的成功率尚未得到充分记录。热探头或双极电凝已被安全有效地应用于出血性结肠血管瘤。推荐使用这些新型电凝探头替代热活检钳来治疗出血性结肠血管瘤。