A dosimetric intercomparison of electron beams in UK radiotherapy centres.

作者信息

Nisbet A, Thwaites D I

机构信息

Department of Medical Physics and Medical Engineering, University of Edinburgh, Western General Hospital, UK.

出版信息

Phys Med Biol. 1997 Dec;42(12):2393-409. doi: 10.1088/0031-9155/42/12/007.

Abstract

A dosimetry intercomparison has been carried out for all 52 radiotherapy centres in the UK which possess electron treatment facilities. The intercomparison was carried out on one treatment unit in each centre and for three energies across the range of available energies. The position of the depth of maximum dose for a standard field size was independently determined and a subsequent beam calibration made. The factor to convert the reading on a calibrated ionization chamber to absorbed dose in an electron beam is energy dependent, and hence to carry out an independent calibration measurement also requires the beam energy to be determined. In addition a quantitative measure of the difference in the calibration chains between the intercomparison equipment and the host department's field instrument was carried out. In order to provide a follow-up to the initial IPSM national photon intercomparison, a photon beam calibration was measured in one photon beam in each centre. For 156 electron beam measurements, a mean ratio of intercomparison measured dose to locally measured dose of 0.994 was obtained with a standard deviation of 1.8%. For the 52 photon beam measurements, a mean ratio of intercomparison measured dose to locally measured dose of 1.003 was obtained with a standard deviation of 1.0%.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验