Validity of pulse oximetry during exercise in elite endurance athletes.

作者信息

Martin D, Powers S, Cicale M, Collop N, Huang D, Criswell D

机构信息

Department of Physical Therapy, University of Florida, Gainesville 32610-0154.

出版信息

J Appl Physiol (1985). 1992 Feb;72(2):455-8. doi: 10.1152/jappl.1992.72.2.455.

Abstract

Eleven highly trained male cyclists [maximal aerobic power (VO2max) = 70.6 +/- 4.2 ml.kg-1.min-1] performed both high intensity constant load (90-95% VO2max) and incremental cycle exercise tests with arterial blood sampling to evaluate the accuracy of pulse oximeter estimates (%SpO2) of arterial oxyhemoglobin fraction of total hemoglobin (%HbO2). Three subjects also performed an incremental exercise test in hypoxic conditions (inspired partial pressure of O2 = 89, 93, or 100 Torr). Arterial %HbO2 was determined via CO-oximetry and ranged from 72 to 99%. Three Ohmeda 3740 pulse oximeters were used to estimate %HbO2, one on each ear lobe and a finger probe. The finger probe tended to provide the best estimate of %HbO2 during exercise: the mean %SpO2 - %HbO2 difference for 232 exercise observations was 0.52 +/- 1.36% (SD). Finger probe %SpO2 and %HbO2 were highly correlated [r = 0.98, standard error of the estimate (SEE) = 1.32%, P less than 0.0001]. The accuracy of pulse oximeters has been questioned during high-intensity exercise. When aerobic power was greater than 81% of VO2max (n = 75), the finger probe's mean error was -0.01 +/- 1.40%. Finger probe %SpO2 and %HbO2 were highly correlated (r = 0.97, SEE = 1.32%, P less than 0.0001). These results indicate that this pulse oximeter is a valid predictor of %HbO2 in elite athletes during cycle exercise.

摘要

11名训练有素的男性自行车运动员[最大有氧功率(VO2max)=70.6±4.2毫升·千克-1·分钟-1]进行了高强度恒定负荷(90-95%VO2max)和递增式自行车运动测试,并采集动脉血样,以评估脉搏血氧仪对总血红蛋白中动脉氧合血红蛋白分数(%HbO2)估计值(%SpO2)的准确性。3名受试者还在低氧条件下(吸入氧分压=89、93或100托)进行了递增运动测试。通过一氧化碳血氧测定法测定动脉%HbO2,范围为72%至99%。使用三台Ohmeda 3740脉搏血氧仪估计%HbO2,每个耳垂上各戴一个,另一个为手指探头。在运动过程中,手指探头往往能提供对%HbO2的最佳估计:232次运动观察的平均%SpO2-%HbO2差值为0.52±1.36%(标准差)。手指探头的%SpO2和%HbO2高度相关[r=0.98,估计标准误差(SEE)=1.32%,P<0.0001]。在高强度运动期间,脉搏血氧仪的准确性受到质疑。当有氧功率大于VO2max的81%(n=75)时,手指探头的平均误差为-0.01±1.40%。手指探头的%SpO2和%HbO2高度相关(r=0.97,SEE=1.32%,P<0.0001)。这些结果表明,这种脉搏血氧仪是精英运动员在自行车运动期间%HbO2的有效预测指标。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验