Luce David A
Reichert Inc., Depew, New York, USA.
J Cataract Refract Surg. 2005 Jan;31(1):156-62. doi: 10.1016/j.jcrs.2004.10.044.
To study the results of an ocular response analyzer (ORA) to determine the biomechanical properties of the cornea and their relationship to intraocular pressure (IOP).
Reichert Inc., Depew, New York, USA.
The ORA (Reichert) makes 2 essentially instantaneous applanation measurements that permit determination of corneal and IOP effects.
Measurements of several populations indicate that corneal hysteresis, a biomechanical measure, varied over a dynamic range of 1.8 to 14.6 mm Hg and was only weakly correlated with corneal thickness (r(2)=0.12); this is related to the observation that some subjects with relatively thick corneas have less-than-average corneal hysteresis. Corneal hysteresis changes diurnally, presumably as a result of hydration changes. Keratoconus, Fuchs' dystrophy, and post-LASIK patients demonstrated low corneal hysteresis.
The corneal hysteresis biomechanical measure may prove valuable for qualification and predictions of outcomes of refractive surgery and in other cases in which corneal biomechanics are important.
研究眼反应分析仪(ORA)测定角膜生物力学特性及其与眼压(IOP)关系的结果。
美国纽约州迪皮尤的Reichert公司。
ORA(Reichert)进行两次基本瞬时的压平测量,从而能够确定角膜和眼压的影响。
对多个群体的测量表明,角膜滞后这一生物力学指标在1.8至14.6毫米汞柱的动态范围内变化,且与角膜厚度仅呈弱相关(r² = 0.12);这与以下观察结果有关,即一些角膜相对较厚的受试者角膜滞后低于平均水平。角膜滞后会随昼夜变化,推测是水化变化所致。圆锥角膜、富克斯营养不良症患者以及准分子激光原位角膜磨镶术(LASIK)术后患者的角膜滞后较低。
角膜滞后这一生物力学指标可能在屈光手术结果的评估和预测以及角膜生物力学起重要作用的其他情况中具有重要价值。