Rodríguez-Galietero Antonio, Montés-Micó Robert, Muñoz Gonzalo, Albarrán-Diego Cesar
Refractive Surgery Unit, Hospital NISA Valencia al Mar, Valencia, Spain.
J Cataract Refract Surg. 2005 Sep;31(9):1736-40. doi: 10.1016/j.jcrs.2005.02.039.
To compare contrast sensitivity and color vision in patients in whom blue-light filtering and non-yellow-tinted intraocular lenses (IOLs) were implanted.
Refractive Surgery Unit, Hospital NISA Valencia al Mar, Valencia, Spain.
Forty eyes of 20 patients were enrolled in a blue-light filtering fellow-eye control study; patients were implanted with a yellow-tinted IOL (AcrySof Natural, Alcon) in 1 eye and a non-yellow-tinted IOL (AcrySof SA60AT, Alcon) in the fellow eye after cataract surgery. Three months postoperatively, monocular contrast sensitivity function was measured with the CSV 1000-E contrast sensitivity chart at distance and the color discrimination with the Farnsworth-Munsell 100 Hue test.
Eyes implanted with blue-light filtering IOLs showed similar contrast sensitivity to that in fellow eyes implanted with non-yellow-tinted IOLs (P>.1). Both types of IOLs showed normal contrast sensitivity values (normalized log-contrast sensitivity about 1.0). There were no statistically significant differences in chromatic discrimination between the 2 types of IOLs (P = .56).
The use of blue-light filtering IOLs is more advisable because they are capable of protecting the retina against ultraviolet light without disturbance of contrast sensitivity and chromatic vision, which produces subjective impairment in visual function.
比较植入蓝光滤过型和非黄色染色人工晶状体(IOL)患者的对比敏感度和色觉。
西班牙巴伦西亚市滨海尼萨医院屈光手术科。
20例患者的40只眼纳入蓝光滤过型对侧眼对照研究;白内障手术后,患者一只眼植入黄色染色IOL(爱尔康公司的AcrySof Natural),对侧眼植入非黄色染色IOL(爱尔康公司的AcrySof SA60AT)。术后3个月,使用CSV 1000-E对比敏感度图表测量远距离单眼对比敏感度功能,并使用 Farnsworth-Munsell 100 色相试验测量色觉辨别能力。
植入蓝光滤过型IOL的眼与植入非黄色染色IOL的对侧眼对比敏感度相似(P>0.1)。两种类型的IOL均显示正常的对比敏感度值(归一化对数对比敏感度约为1.0)。两种类型的IOL在色觉辨别方面无统计学显著差异(P = 0.56)。
使用蓝光滤过型IOL更可取,因为它们能够保护视网膜免受紫外线伤害,同时不干扰对比敏感度和色觉,而色觉会导致视觉功能的主观损害。