用于非均匀弹性成像体模的组织模拟琼脂/明胶材料。

Tissue-mimicking agar/gelatin materials for use in heterogeneous elastography phantoms.

作者信息

Madsen Ernest L, Hobson Maritza A, Shi Hairong, Varghese Tomy, Frank Gary R

机构信息

Department of Medical Physics, University of Wisconsin, Madison, WI 53706, USA.

出版信息

Phys Med Biol. 2005 Dec 7;50(23):5597-618. doi: 10.1088/0031-9155/50/23/013. Epub 2005 Nov 16.

Abstract

Five 9 cm x 9 cm x 9 cm phantoms, each with a 2-cm-diameter cylindrical inclusion, were produced with various dry-weight concentrations of agar and gelatin. Elastic contrasts ranged from 1.5 to 4.6, and values of the storage modulus (real part of the complex Young's modulus) were all in the soft tissue range. Additives assured immunity from bacterial invasion and can produce tissue-mimicking ultrasound and NMR properties. Monitoring of strain ratios over a 7 to 10 month period indicated that the mechanical properties of the phantoms were stable, allowing about 1 month for the phantom to reach chemical equilibrium. The only dependable method for determining the storage moduli of the inclusions is to make measurements on samples excised from the phantoms. If it is desired to produce and accurately characterize a phantom with small inclusions with other shapes, such as an array of small spheres, an auxiliary phantom with the geometry of the cylindrical inclusion phantoms or the equivalent should be made at the same time using the same materials. The elastic contrast can then be determined using samples excised from the auxiliary phantom. A small increase of about 10% in volume of the cylindrical inclusions occurred-a tolerable increase. Interestingly, the smallest increase (about 5%) occurred in the phantom with the largest elastic contrast.

摘要

制作了5个9厘米×9厘米×9厘米的体模,每个体模都有一个直径2厘米的圆柱形内含物,这些体模由不同干重浓度的琼脂和明胶制成。弹性对比度在1.5至4.6之间,储能模量(复杨氏模量的实部)值均在软组织范围内。添加剂确保了体模不受细菌侵袭,并能产生类似组织的超声和核磁共振特性。在7至10个月的时间内对应变比进行监测表明,体模的力学性能是稳定的,体模达到化学平衡大约需要1个月时间。确定内含物体模储能模量的唯一可靠方法是对从体模上切下的样本进行测量。如果希望制作并准确表征带有其他形状小内含物(如小球体阵列)的体模,应同时使用相同材料制作一个具有圆柱形内含物体模几何形状或等效几何形状的辅助体模。然后可以使用从辅助体模上切下的样本确定弹性对比度。圆柱形内含物体积出现了约10%的小幅增加——这是可以接受的增加幅度。有趣的是,弹性对比度最大的体模中增加幅度最小(约5%)。

相似文献

1
Tissue-mimicking agar/gelatin materials for use in heterogeneous elastography phantoms.
Phys Med Biol. 2005 Dec 7;50(23):5597-618. doi: 10.1088/0031-9155/50/23/013. Epub 2005 Nov 16.
2
Stability of heterogeneous elastography phantoms made from oil dispersions in aqueous gels.
Ultrasound Med Biol. 2006 Feb;32(2):261-70. doi: 10.1016/j.ultrasmedbio.2005.10.009.
3
Characterization of biomechanical properties of agar based tissue mimicking phantoms for ultrasound stiffness imaging techniques.
J Mech Behav Biomed Mater. 2014 Jul;35:132-43. doi: 10.1016/j.jmbbm.2014.03.017. Epub 2014 Apr 1.
4
Spherical lesion phantoms for testing the performance of elastography systems.
Phys Med Biol. 2005 Dec 21;50(24):5983-95. doi: 10.1088/0031-9155/50/24/015. Epub 2005 Dec 6.
5
Tissue-mimicking oil-in-gelatin dispersions for use in heterogeneous elastography phantoms.
Ultrason Imaging. 2003 Jan;25(1):17-38. doi: 10.1177/016173460302500102.
6
Development of oil-in-gelatin phantoms for viscoelasticity measurement in ultrasound shear wave elastography.
Ultrasound Med Biol. 2014 Jan;40(1):168-76. doi: 10.1016/j.ultrasmedbio.2013.08.020. Epub 2013 Oct 18.
7
Copolymer-in-oil phantom materials for elastography.
Ultrasound Med Biol. 2009 Jul;35(7):1185-97. doi: 10.1016/j.ultrasmedbio.2009.01.012. Epub 2009 May 7.
9
Anthropomorphic breast phantoms for testing elastography systems.
Ultrasound Med Biol. 2006 Jun;32(6):857-74. doi: 10.1016/j.ultrasmedbio.2006.02.1428.

引用本文的文献

1
Development of breast-mimicking phantoms for use in optical coherence elastography.
J Biomed Opt. 2025 Dec;30(12):124504. doi: 10.1117/1.JBO.30.12.124504. Epub 2025 Sep 2.
2
Lipid Microparticle-Based Phantoms Modeling Hepatic Steatosis for the Validation of Quantitative Imaging Techniques.
Small Methods. 2025 Aug;9(8):e2500043. doi: 10.1002/smtd.202500043. Epub 2025 Apr 25.
3
Development of breast tissue-mimicking electrical and acoustic phantoms for magneto-acoustic electrical tomography.
Ann N Y Acad Sci. 2025 May;1547(1):192-203. doi: 10.1111/nyas.15338. Epub 2025 Apr 15.
7
A first-in-human study of quantitative ultrasound to assess transplant kidney fibrosis.
Nat Med. 2025 Mar;31(3):970-978. doi: 10.1038/s41591-024-03417-5. Epub 2025 Mar 3.
8
3D-printed neck phantoms with detailed anatomy for ultrasound-guided procedure and device testing.
Laryngoscope Investig Otolaryngol. 2024 Aug 6;9(4):e1309. doi: 10.1002/lio2.1309. eCollection 2024 Aug.
9
Agarose as a Tissue Mimic for the Porcine Heart, Kidney, and Liver: Measurements and a Springpot Model.
Bioengineering (Basel). 2024 Jun 8;11(6):589. doi: 10.3390/bioengineering11060589.
10
A Dynamic Ultrasound Phantom with Tissue-Mimicking Mechanical and Acoustic Properties.
Adv Sci (Weinh). 2024 Jun;11(22):e2400271. doi: 10.1002/advs.202400271. Epub 2024 Apr 22.

本文引用的文献

1
Spherical lesion phantoms for testing the performance of elastography systems.
Phys Med Biol. 2005 Dec 21;50(24):5983-95. doi: 10.1088/0031-9155/50/24/015. Epub 2005 Dec 6.
2
Tissue-mimicking oil-in-gelatin dispersions for use in heterogeneous elastography phantoms.
Ultrason Imaging. 2003 Jan;25(1):17-38. doi: 10.1177/016173460302500102.
3
Contrast-transfer efficiency for continuously varying tissue moduli: simulation and phantom validation.
Ultrasound Med Biol. 2001 Aug;27(8):1115-25. doi: 10.1016/s0301-5629(01)00411-2.
5
Interlaboratory comparison of ultrasonic backscatter, attenuation, and speed measurements.
J Ultrasound Med. 1999 Sep;18(9):615-31. doi: 10.7863/jum.1999.18.9.615.
6
Anthropomorphic 1H MRS head phantom.
Med Phys. 1998 Jul;25(7 Pt 1):1145-56. doi: 10.1118/1.598306.
7
Elastic and acoustic properties of vessel mimicking material for elasticity imaging.
Ultrason Imaging. 1997 Apr;19(2):112-26. doi: 10.1177/016173469701900202.
8
Sonoelasticity imaging: theory and experimental verification.
J Acoust Soc Am. 1995 Jun;97(6):3875-86. doi: 10.1121/1.412399.
9
Low-contrast focal lesion detectability phantom for 1H MR imaging.
Med Phys. 1991 May-Jun;18(3):549-54. doi: 10.1118/1.596659.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验