采用外源性气体重呼吸法无创测定心脏病患者心输出量的验证

Validation of the use of foreign gas rebreathing method for non-invasive determination of cardiac output in heart disease patients.

作者信息

Dong Liang, Wang Jian-an, Jiang Chen-yang

机构信息

Department of Cardiology, Sir Run Run Shaw Hospital, School of Medicine, Zhejiang University, Hangzhou 310016, China.

出版信息

J Zhejiang Univ Sci B. 2005 Dec;6(12):1157-62. doi: 10.1631/jzus.2005.B1157.

Abstract

OBJECTIVE

To compare a new device (Innocor) for non-invasive measurement of cardiac output (CO) by foreign gas rebreathing method with conventional techniques used in the measurements of cardiac function.

METHODS

Cardiac outputs measured by Innocor (CO(RB)) were compared with CO obtained by echocardiography (CO(EC)), Swan-Ganz thermodilution (CO(TD)), and left ventricle radiography (CO(LVR)) in 34 patients subjected to cardiac catheterization. Values obtained from the four methods were analyzed by linear regression and paired values were compared by the method of Bland and Altman in SPSS.

RESULTS

There was strong positive correlation (r=0.94) between Innocor cardiac output values and the corresponding values obtained by thermodilution and between CO(EC) and CO(LVR) values. Thermodilution appears to overestimate cardiac output when compared to the values obtained with Innocor by (0.66+/-0.22) L/min (P<0.0001). There was no correlation between data obtained by Innocor and the corresponding CO(EC) and CO(LVR) values.

CONCLUSION

Innocor CO(RB) is an easy, safe and well established method for non-invasive measurement of cardiac output with good prospects for clinical application in heart disease patients.

摘要

目的

采用外源性气体重呼吸法,使用一种新型设备(Innocor)无创测量心输出量(CO),并与测量心功能的传统技术进行比较。

方法

在34例接受心导管检查的患者中,将Innocor测量的心输出量(CO(RB))与经超声心动图(CO(EC))、Swan-Ganz热稀释法(CO(TD))和左心室造影术(CO(LVR))获得的心输出量进行比较。采用线性回归分析这四种方法获得的值,并在SPSS中使用Bland和Altman方法比较配对值。

结果

Innocor心输出量值与热稀释法获得的相应值之间以及CO(EC)和CO(LVR)值之间存在强正相关(r = 0.94)。与Innocor获得的值相比,热稀释法似乎高估了心输出量,差值为(0.66±0.22)L/min(P < 0.0001)。Innocor获得的数据与相应的CO(EC)和CO(LVR)值之间无相关性。

结论

Innocor CO(RB)是一种简便、安全且成熟的无创测量心输出量的方法,在心脏病患者中具有良好的临床应用前景。

相似文献

引用本文的文献

8
Non-Invasive Monitoring of Cardiac Output in Critical Care Medicine.重症医学中的心输出量无创监测
Front Med (Lausanne). 2017 Nov 20;4:200. doi: 10.3389/fmed.2017.00200. eCollection 2017.
9
Cardiac function in adults following minimally invasive repair of pectus excavatum.漏斗胸微创修复术后成人的心脏功能
Interact Cardiovasc Thorac Surg. 2016 May;22(5):525-9. doi: 10.1093/icvts/ivw007. Epub 2016 Feb 8.

本文引用的文献

5
Effect of ventilation and perfusion imbalance on inert gas rebreathing variables.通气与灌注失衡对惰性气体再呼吸变量的影响。
J Appl Physiol Respir Environ Exerc Physiol. 1984 Feb;56(2):364-9. doi: 10.1152/jappl.1984.56.2.364.
9
Lung tissue volume and blood flow by rebreathing theory.基于重复呼吸理论的肺组织容积与血流情况
J Appl Physiol Respir Environ Exerc Physiol. 1978 May;44(5):795-802. doi: 10.1152/jappl.1978.44.5.795.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验