Abstract

Implantable cardioverter-defibrillator (ICD) shocks diminish patients' quality of life, increase health care resource utilization, and may lead to other adverse sequelae. Better understanding of the factors that lead to ICD therapies, and better strategies to avoid unnecessary therapies, are needed to optimize patient outcomes. Data from major randomized clinical trials involving the use of ICDs and cardiac resynchronization therapy-defibrillator devices were reviewed to determine control group mortality rates, control group sudden death rates, and the frequency of appropriate and inappropriate ICD therapies. In all studies that classified deaths, appropriate ICD therapies outnumbered control group sudden cardiac deaths by a factor of 2 to 3. Some of these episodes can be explained by device programming, by the treatment of potentially unsustained tachycardias, and by errors of episode classification. Another underexplored possibility is that device therapy is proarrhythmic. Reasons for frequent therapies and methods to prevent them are discussed, as well as the notion of device proarrhythmia and the potentially detrimental effects of ICD shocks. These issues clearly affect the overall benefit of device therapy and have important implications for patient management and health care delivery.

摘要

植入式心脏复律除颤器(ICD)电击会降低患者的生活质量,增加医疗资源的使用,并可能导致其他不良后果。为了优化患者的治疗效果,需要更好地了解导致ICD治疗的因素,以及避免不必要治疗的更好策略。回顾了涉及使用ICD和心脏再同步治疗除颤器设备的主要随机临床试验数据,以确定对照组的死亡率、对照组的心源性猝死率以及适当和不适当的ICD治疗频率。在所有对死亡进行分类的研究中,适当的ICD治疗数量比对照组的心源性猝死数量多2至3倍。其中一些情况可以通过设备编程、对潜在非持续性心动过速的治疗以及事件分类错误来解释。另一个尚未充分探索的可能性是设备治疗具有促心律失常作用。讨论了频繁治疗的原因和预防方法,以及设备促心律失常的概念和ICD电击的潜在有害影响。这些问题显然会影响设备治疗的总体益处,并对患者管理和医疗服务提供具有重要意义。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验