Suppr超能文献

手持测力计对老年人股四头肌力量的测量有用吗?与金标准Bodex测力计的比较。

Is hand-held dynamometry useful for the measurement of quadriceps strength in older people? A comparison with the gold standard Bodex dynamometry.

作者信息

Martin H J, Yule V, Syddall H E, Dennison E M, Cooper C, Aihie Sayer A

机构信息

MRC Epidemiology Resource Centre, University of Southampton, Southampton, UK.

出版信息

Gerontology. 2006;52(3):154-9. doi: 10.1159/000091824.

Abstract

BACKGROUND

The lower limb muscle strength is an important determinant of physical function in older people. However, measurement in clinical and epidemiological settings has been limited because of the requirement for large-scale equipment. A protocol using a novel, versatile hand-held dynamometer (HHD) has been developed to measure the quadriceps strength in a supine position.

OBJECTIVE

The objective of this study was to assess the validity of this new methodology for measuring the lower limb muscle strength compared to the gold standard Biodex dynamometer.

METHODS

The supine quadriceps strength was measured twice with each of the Biodex and the HHD in 20 men and women, aged 61-81 years, on their non-dominant leg. The agreement between the peak torques obtained by Biodex and HHD was analyzed.

RESULTS

The mean peak Biodex and HHD results were 83.4 +/- (SD) 28.0 Nm and 68.9 +/- 19.6 Nm, respectively. The HHD undermeasured the quadriceps strength by an average of 14.5 Nm (95% CI 8.5, 20.6) compared to the Biodex, and this effect was most marked in the strongest participants. Nevertheless, there was a good correlation between the measures (r = 0.91, p < 0.0001). Classification of individuals into tertiles of muscle strength showed good agreement between the two methods (Kappa = 0.69, p < 0.0001).

CONCLUSIONS

Our findings suggest that the HHD using a supine positioning offers a feasible, inexpensive, and portable test of quadriceps muscle strength for use in healthy older people. It underestimates the absolute quadriceps strength compared to the Biodex particularly in stronger people, but is a useful tool for ranking muscle strength of older people in epidemiological studies. It may also be of value for quick and objective assessment of physical function in the clinical setting.

摘要

背景

下肢肌肉力量是老年人身体功能的重要决定因素。然而,由于需要大型设备,临床和流行病学环境中的测量受到限制。已开发出一种使用新型多功能手持测力计(HHD)的方案,用于测量仰卧位时的股四头肌力量。

目的

本研究的目的是评估与金标准Biodex测力计相比,这种测量下肢肌肉力量的新方法的有效性。

方法

对20名年龄在61 - 81岁的男性和女性的非优势腿,分别使用Biodex和HHD测量仰卧位股四头肌力量两次。分析Biodex和HHD获得的峰值扭矩之间的一致性。

结果

Biodex和HHD的平均峰值结果分别为83.4±(标准差)28.0 Nm和68.9±19.6 Nm。与Biodex相比,HHD测量的股四头肌力量平均低估了14.5 Nm(95%可信区间8.5, 20.)这种影响在最强壮的参与者中最为明显。然而,两种测量方法之间存在良好的相关性(r = 0.91,p < 0.0001)。将个体按肌肉力量三分位数分类显示两种方法之间具有良好的一致性(Kappa = 0.69,p < 0.0001)。

结论

我们的研究结果表明,使用仰卧位的HHD为健康老年人提供了一种可行、廉价且便携的股四头肌力量测试方法。与Biodex相比,它低估了绝对股四头肌力量,尤其是在较强壮的人群中,但它是流行病学研究中对老年人肌肉力量进行排名的有用工具。它在临床环境中对身体功能的快速客观评估中也可能具有价值。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验