Jiang Steve B
Department of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, MA 02114, USA.
Med Dosim. 2006 Summer;31(2):141-51. doi: 10.1016/j.meddos.2005.12.005.
In this review article, we discuss various technical aspects of image-guided respiration-gated radiation therapy. We first review some basic concepts related to respiratory gating, including gating window, duty cycle, residual motion, internal/external gating, amplitude/phase gating, etc. We then discuss 2 implementations of image-guided respiration-gated treatment, i.e., the Mitsubishi/Hokkaido technique for internal gating and the MGH technique for external gating. Several existing problems related to respiratory gating, namely external gating mode (phase vs. amplitude), imaging dose for internal gating, gated treatment for lung cancer without implanted fiducial makers, as well as gated intensity-modulated radiation therapy issues, are also discussed along with potential solutions.
在这篇综述文章中,我们讨论了图像引导呼吸门控放射治疗的各种技术方面。我们首先回顾一些与呼吸门控相关的基本概念,包括门控窗口、占空比、残余运动、内/外门控、幅度/相位门控等。然后我们讨论图像引导呼吸门控治疗的两种实施方式,即用于内门控的三菱/北海道技术和用于外门控的MGH技术。还讨论了与呼吸门控相关的几个现有问题,即外门控模式(相位与幅度)、内门控的成像剂量、无植入基准标记物的肺癌门控治疗以及门控调强放射治疗问题,并提出了潜在的解决方案。