一种新型乳腺双能X线吸收测定技术与乳腺X线密度的临床比较。

Clinical comparison of a novel breast DXA technique to mammographic density.

作者信息

Shepherd John A, Herve Lionel, Landau Jessie, Fan Bo, Kerlikowske Karla, Cummings Steve R

机构信息

Musculoskeletal and Quantitative Imaging Research Group, Department of Radiology, University of California, San Francisco, California 94143, USA.

出版信息

Med Phys. 2006 May;33(5):1490-8. doi: 10.1118/1.2193691.

Abstract

We compare mammography breast density (BD(MD)) to the measure of breast composition using a clinical dual energy absorptiometry (DXA) system (BD(DXA)) calibrated to measure breast density. A DXA scanning protocol was developed to scan breasts isolated in the DXA scan field in either a prone pendulous or decubitus mediolateral position. A total of 17 participants were recruited among women undergoing clinical mammography examinations. Each participant had duplicate DXA scans and duplicate craniocaudal-view mammograms of their right breast with repositioning between each scan and one DXA and one craniocaudal-view mammogram of their left breast. The in vivo repeatability (RMS SD) of BD(DXA) and BD(MD) on duplicate scans was found to be 1.2% for BD(DXA) and 1.4% for BD(MD) when repeat BD(MD) measures were made on the same day. When repeat BD(MD) measures of the same breast were made more than 50 days apart, the repeatability decreased to 5.5%. Left and right breast measurements were highly correlated with both techniques at r2 = 0.98 for BD(DXA) and r2 = 0.86 for BD(MD). Moderate correlation (r2 = 0.52) was found between BD(DXA) and BD(MD) measurements. However, after recalibrating the DXA system to mammography reference materials, negative percent fibroglandular values were measured for the most fatty breasts. Thus, our results are reproducible and accurate to common mammography tissue standards, but did not accurately reflect true percent fibroglandular levels and further development of phantom standards are necessary. We conclude that breast composition can be precisely evaluated and assessed with clinical DXA densitometers at a lower dose than with mammographic breast density methods.

摘要

我们将乳腺X线摄影的乳房密度(BD(MD))与使用临床双能X线吸收测定法(DXA)系统测量乳房组成的方法(BD(DXA))进行比较,该DXA系统经过校准以测量乳房密度。制定了一种DXA扫描方案,用于扫描在DXA扫描视野中处于俯卧下垂或侧卧中外侧位的孤立乳房。在接受临床乳腺X线摄影检查的女性中总共招募了17名参与者。每位参与者对其右乳房进行了两次重复的DXA扫描和两次重复的头尾位乳腺X线摄影,每次扫描之间进行重新定位,对其左乳房进行了一次DXA扫描和一次头尾位乳腺X线摄影。当在同一天进行重复的BD(MD)测量时,BD(DXA)和BD(MD)在重复扫描上的体内重复性(均方根标准差)对于BD(DXA)为1.2%,对于BD(MD)为1.4%。当对同一乳房的BD(MD)测量相隔超过50天时,重复性降至5.5%。两种技术对左右乳房的测量高度相关,BD(DXA)的r2 = 0.98,BD(MD)的r2 = 0.86。BD(DXA)和BD(MD)测量之间发现中度相关性(r2 = 0.52)。然而,在将DXA系统重新校准至乳腺X线摄影参考材料后,对于脂肪含量最高的乳房测量到了负的纤维腺体百分比值。因此,我们的结果对于常见的乳腺X线摄影组织标准是可重复且准确的,但并未准确反映真实的纤维腺体百分比水平,需要进一步开发体模标准。我们得出结论,与乳腺X线摄影乳房密度方法相比,使用临床DXA密度计可以以更低剂量精确评估和评价乳房组成。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验