An intercomparison of 11 amorphous silicon EPIDs of the same type: implications for portal dosimetry.

作者信息

Winkler Peter, Georg Dietmar

机构信息

Division of Medical Radiation Physics, Department of Radiotherapy and Radiobiology, Medical University of Vienna, Waehringer Guertel 18-20, A-1090 Vienna, Austria.

出版信息

Phys Med Biol. 2006 Sep 7;51(17):4189-200. doi: 10.1088/0031-9155/51/17/005. Epub 2006 Aug 8.

Abstract

The use of electronic portal imaging devices (EPIDs) for portal dosimetry requires knowledge of their dosimetric properties. The pixel value response of amorphous silicon EPIDs of type Elekta iViewGT is known to be nonlinear with dose. However, it is not clear whether these nonlinearities vary with time and from one detector to another, respectively. In the present study, the dose-response characteristics of 11 iViewGT EPIDs were investigated with respect to dose rate, total dose and field size. It was found that each detector needs to be individually calibrated, not only in terms of absolute sensitivity but also with respect to its relative response variations with exposure parameters. Doubling the dose rate typically increased the EPID signal between 1.4% and 2.8%. Changing the number of monitor units from 30 to 500 was accompanied by an increase in detector sensitivity between 1.7% and 2.8%. The EPID scatter factors were always within +/-1%. It was observed that the dose-response behaviour was not stable with respect to time. Particularly within the first weeks of operation, detector ageing caused variations in both absolute sensitivity and relative response curves. It is recommended to establish a quality assurance programme if the amorphous silicon EPIDs are intended to be used for clinical portal dosimetry.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验