Boswell Sarah, Tomé Wolfgang, Jeraj Robert, Jaradat Hazim, Mackie T Rock
Department of Medical Physics, University of Wisconsin, Madison, Wisconsin 53706, USA.
Med Phys. 2006 Nov;33(11):4395-404. doi: 10.1118/1.2349698.
Precise daily target localization is necessary to achieve highly conformal radiation delivery. In helical tomotherapy, setup verification may be accomplished just prior to delivering each fraction by acquiring a megavoltage CT scan of the patient in the treatment position. This daily image set may be manually or automatically registered to the image set on which the treatment plan was calculated, in order to determine any needed adjustments. The system was tested by acquiring 104 MVCT scans of an anthropomorphic head phantom to which translational displacements had been introduced with respect to the planning image set. Registration results were compared against an independent, optically guided positioning system. The total experimental uncertainty was within approximately 1 mm. Although the registration of phantom images is not fully analogous to the registration of patient images, this study confirms that the system is capable of phantom localization with sub-voxel accuracy. In seven registration problems considered, expert human observers were able to perform manual registrations with comparable or inferior accuracy to automatic registration by mutual information. The time to compute an automatic registration is considerably shorter than the time required for manual registration. However, human evaluation of automatic results is necessary in order to identify occasional outliers, and to ensure that the registration is clinically acceptable, especially in the case of deformable patient anatomy.
精确的每日靶区定位对于实现高度适形放疗至关重要。在螺旋断层放疗中,可在每次放疗前通过获取处于治疗体位的患者的兆伏级CT扫描来完成摆位验证。该每日图像集可手动或自动与计算治疗计划时所用的图像集配准,以确定任何必要的调整。通过获取对其相对于计划图像集引入了平移位移的仿真人头模的104次兆伏级CT扫描来测试该系统。将配准结果与独立的光学引导定位系统进行比较。总的实验不确定性在约1毫米以内。尽管模体图像的配准与患者图像的配准并不完全类似,但本研究证实该系统能够以亚体素精度进行模体定位。在所考虑的七个配准问题中,专业的人类观察者能够进行手动配准,其精度与通过互信息进行的自动配准相当或更低。计算自动配准的时间比手动配准所需的时间短得多。然而,为了识别偶尔出现的异常值,并确保配准在临床上是可接受的,尤其是在患者解剖结构可变形的情况下,需要对自动结果进行人工评估。