J Glaucoma. 2007 Jan;16(1):29-35. doi: 10.1097/IJG.0b013e31802b350f.
To determine whether there is a temporal relationship between the diurnal variation of central corneal thickness (CCT) and intraocular pressure (IOP) by Goldmann applanation tonometry in young normal human participants in vivo.
Twenty-five eyes of 25 young healthy normal participants were examined in a prospective observational cross-sectional study. IOP, CCT and corneal curvature were measured using standard clinical techniques over a 24-hour period, and the temporal interrelationships between these parameters were examined.
The overnight change in IOP measured by Goldmann tonometry was 3.1+/-2.4 mm Hg (P<0.001), CCT was 20.1+/-10.9 mum (P=0.016), with no statistical change in central corneal curvature (0.05 mm, P=0.477, paired t test with Bonferroni correction). Both IOP and CCT were highest on awakening at 7:00 then dropped rapidly to baseline levels by 9:00 (linear mixed models), and these two parameters were highly correlated (r=0.978, P<0.001). After 9:00, there was no correlation between these parameters (r=-0.453, P=0.260).
The results of this study have highlighted a potential link between the diurnal variation of CCT and the accuracy of Goldmann tonometry estimates of IOP during the first 2 hours after awakening. Clinicians should be wary of using Goldmann tonometry to estimate IOP until the overnight increase in CCT has resolved.
通过Goldmann压平眼压计测定年轻正常人类受试者活体中央角膜厚度(CCT)与眼压(IOP)的昼夜变化之间是否存在时间关系。
在一项前瞻性观察性横断面研究中,对25名年轻健康正常受试者的25只眼睛进行了检查。在24小时内使用标准临床技术测量眼压、中央角膜厚度和角膜曲率,并检查这些参数之间的时间相互关系。
通过Goldmann眼压计测量的眼压夜间变化为3.1±2.4 mmHg(P<0.001),中央角膜厚度为20.1±10.9μm(P=0.016),中央角膜曲率无统计学变化(0.05 mm,P=0.477,采用Bonferroni校正的配对t检验)。眼压和中央角膜厚度在早上7点醒来时最高,然后在9点迅速降至基线水平(线性混合模型),这两个参数高度相关(r=0.978,P<0.001)。9点之后,这些参数之间无相关性(r=-0.453,P=0.260)。
本研究结果突出了中央角膜厚度的昼夜变化与觉醒后最初2小时内Goldmann眼压计眼压估计准确性之间的潜在联系。在中央角膜厚度的夜间增加消退之前,临床医生在使用Goldmann眼压计估计眼压时应谨慎。