Hougaard Jesper Leth, Heijl Anders, Bengtsson Boel
Department of Clinical Sciences, Ophthalmology, Malmö University Hospital, Lund University, Sweden.
J Glaucoma. 2007 May;16(3):302-6. doi: 10.1097/IJG.0b013e318032e4d4.
To investigate the diagnostic accuracy of retinal nerve fiber layer thickness (RNFLT) summary report parameters from Stratus optical coherence tomography (OCT) in glaucoma patients.
We obtained Stratus OCT measurements of peripapillary circle scans (average values) of the regular image resolution "FAST RNFLT" protocol, and of 1 circle scan of the high resolution "RNFLT" protocol in one eye of each of 62 glaucoma patients with mild or moderate visual field (VF) loss and 90 healthy subjects. Sensitivity, specificity, and diagnostic accuracy [(true positive+true negative)/all] were evaluated for all summary report parameters including the newer (eg, "Imax," the maximum thickness point in the inferior quadrant) at the normative limits of the Stratus OCT.
The diagnostic accuracy of full circle RNFLT using the 5% normal limit was 89% with the FAST RNFLT and 87% with the RNFLT protocol; this was at least as good as any other parameter. The diagnostic performance of the 2 protocols did not differ significantly for most parameters. In eyes with mild VF loss (n=39) diagnostic sensitivities reached 72% and 77% at specificities >or=95% using the FAST RNFLT and RNFLT protocol, respectively.
The diagnostic accuracy of the full circle RNFLT was as good as any Stratus OCT parameter on the basis of the peripapillary RNFL thickness measurements, including localized measurements. The sensitivity was moderately high in patients with mild glaucomatous VF loss. There seems to be room for further development of OCT interpretation tools for early diagnosis of glaucoma.
研究Stratus光学相干断层扫描(OCT)的视网膜神经纤维层厚度(RNFLT)总结报告参数在青光眼患者中的诊断准确性。
我们获取了62例轻度或中度视野(VF)缺损的青光眼患者及90名健康受试者单眼的Stratus OCT测量结果,这些测量结果来自常规图像分辨率“FAST RNFLT”方案的视乳头周围环形扫描(平均值)以及高分辨率“RNFLT”方案的1次环形扫描。针对所有总结报告参数,包括Stratus OCT正常范围的更新参数(如“Imax”,即下象限的最大厚度点),评估其敏感性、特异性及诊断准确性[(真阳性+真阴性)/总数]。
使用5%正常范围时,FAST RNFLT方案全环RNFLT的诊断准确性为89%,RNFLT方案为87%;这至少与其他任何参数一样好。对于大多数参数,这两种方案的诊断性能无显著差异。在轻度VF缺损的眼中(n = 39),使用FAST RNFLT和RNFLT方案时,特异性≥95%时诊断敏感性分别达到72%和77%。
基于视乳头周围RNFL厚度测量,全环RNFLT的诊断准确性与Stratus OCT的任何参数一样好,包括局部测量。在轻度青光眼性VF缺损患者中敏感性中等偏高。对于青光眼的早期诊断,OCT解读工具似乎还有进一步发展的空间。