Rocha Karolinne Maia, Soriano Eduardo S, Chamon Wallace, Chalita Maria Regina, Nosé Walton
Department of Ophthalmology, Paulista School of Medicine, Federal University of São Paulo, São Paulo, Brazil.
Ophthalmology. 2007 Nov;114(11):2050-4. doi: 10.1016/j.ophtha.2007.01.024. Epub 2007 Apr 19.
To compare distance-corrected, near, and intermediate visual acuities as a measurement of depth of focus and spherical aberration of eyes implanted with aspheric and spherical intraocular lenses (IOLs).
Randomized prospective study.
One hundred twenty eyes of 60 patients with bilateral cataract implanted with 3 IOL models (AcrySof IQ, AcrySof SN60AT [Alcon Laboratories, Fort Worth, TX], and Sensar AR40 [Allergan Surgical, Irvine, CA]).
Pupil diameter was analyzed by using a Colvard pupillometer (OASIS Medical, Inc., Glendora, CA) under photopic, mesopic, and scotopic conditions. Distance (6 m), intermediate (1 m), and near (0.33 m) visual acuities were measured with distance correction in place 90 days after surgery. Wavefront analysis (LadarWave; Alcon Laboratories) was performed using 5-mm and measured photopic pupil diameter at 30 and 90 days after surgery.
To evaluate spherical aberration and depth of focus (by means of distance-corrected near and intermediate visual acuity) in patients implanted with aspheric and spherical IOLs.
Photopic pupil sizes in each group were similar at 30 days after surgery. At 90 days after surgery, mean logarithm of the minimum angle of resolution distance-corrected near visual acuity (+/-standard deviation [SD]) was 0.50+/-0.20 in the AcrySof IQ group, 0.38+/-0.17 in the AcrySof SN60AT group, and 0.45+/-0.16 in Sensar AR40 group. Mean spherical aberration values (+/-SD) were 0.03+/-0.05 microm in the AcrySof IQ group, 0.24+/-0.04 microm in the AcrySof SN60AT group, and 0.14+/-0.07 microm in the Sensar AR40 group. The AcrySof IQ group showed a statistically significant lower induction of spherical aberration and worse distance-corrected near visual acuity. The AcrySof SN60AT group showed statistically significant higher mean spherical aberration values and better distance-corrected near and intermediate visual acuity.
The reduction of total spherical aberration after aspheric IOL implantation may degrade distance-corrected near and intermediate visual acuity.
比较经距离矫正的近视力和中视力,以此作为植入非球面和球面人工晶状体(IOL)眼的焦深和球差的一种测量方法。
随机前瞻性研究。
60例双侧白内障患者的120只眼,植入3种IOL模型(AcrySof IQ、AcrySof SN60AT[爱尔康实验室,沃思堡,德克萨斯州]和Sensar AR40[眼力健外科,尔湾,加利福尼亚州])。
使用科尔瓦德瞳孔计(绿洲医疗公司,格伦多拉,加利福尼亚州)在明视、 mesopic和暗视条件下分析瞳孔直径。术后90天,在有距离矫正的情况下测量远视力(6米)、中视力(1米)和近视力(0。33米)。术后30天和90天,使用5毫米测量的明视瞳孔直径进行波前分析(LadarWave;爱尔康实验室)。
评估植入非球面和球面IOL患者的球差和焦深(通过经距离矫正的近视力和中视力)。
术后30天,每组的明视瞳孔大小相似。术后90天,AcrySof IQ组经距离矫正的近视力最小分辨角对数平均值(±标准差[SD])为0。50±0。20,AcrySof SN60AT组为0。38±0。17,Sensar AR40组为0。45±0。16。AcrySof IQ组的平均球差值(±SD)为0。03±0。05微米,AcrySof SN60AT组为0。24±0。04微米,Sensar AR40组为0。14±0。07微米。AcrySof IQ组的球差诱导在统计学上显著较低,经距离矫正的近视力较差。AcrySof SN60AT组的平均球差值在统计学上显著较高,经距离矫正的近视力和中视力较好。
植入非球面IOL后总球差的降低可能会降低经距离矫正的近视力和中视力。