角膜屈光手术后Orbscan II和Pentacam测量中央角膜厚度的比较。

Comparison of central corneal thickness measurements by Orbscan II and Pentacam after corneal refractive surgery.

作者信息

Matsuda Jumpei, Hieda Osamu, Kinoshita Shigeru

机构信息

Department of Ophthalmology, Kyoto Prefectural University of Medicine, Kyoto, Japan.

Department of Ophthalmology, National Center for Geriatrics and Gerontology, Obu, Japan.

出版信息

Jpn J Ophthalmol. 2008 Jul-Aug;52(4):245-249. doi: 10.1007/s10384-008-0550-x. Epub 2008 Sep 5.

Abstract

PURPOSE

To compare the accuracy of postoperative pachymetry between Orbscan II (Bausch & Lomb) scanning-slit corneal topography/pachymetry and the Pentacam (Oculus) rotating Scheimpflug camera.

METHODS

Central corneal thickness (CCT) was determined in 24 patients (48 eyes) before and after laser in situ keratomileusis (LASIK) or Epipolis LASIK (Epi-LASIK) procedures. All eyes were examined by Orbscan II and Pentacam prior to refractive surgery and at the first, fourth, and twelfth week postoperatively. The residual CCT (RCCT) measured by each instrument was compared to the theoretical RCCT.

RESULTS

On the first, fourth, and twelfth week after the refractive surgery, the mean RCCT measurements by Orbscan II were 413 +/- 72, 435 +/- 65, and 440 +/- 69 microm, respectively. Those of Pentacam were 434 +/- 51, 436 +/- 53, and 438 +/- 50 microm, respectively. Orbscan II measurements at the postoperative first and fourth week were significantly smaller than the theoretical RCCT (P < 0.01, P < 0.01, paired t test). There was no statistical difference between the theoretical RCCT and the Pentacam measurements at any stage.

CONCLUSION

The Orbscan II measurement values of postoperative corneas tended to be thinner than the theoretical values, but not those of the Pentacam.

摘要

目的

比较Orbscan II(博士伦)扫描裂隙角膜地形图/测厚仪与Pentacam(欧科路)旋转式Scheimpflug相机术后测厚的准确性。

方法

在24例患者(48只眼)行准分子原位角膜磨镶术(LASIK)或上皮瓣下角膜磨镶术(Epi-LASIK)前后测定中央角膜厚度(CCT)。所有眼睛在屈光手术前以及术后第1周、第4周和第12周均接受Orbscan II和Pentacam检查。将每种仪器测量的残余中央角膜厚度(RCCT)与理论RCCT进行比较。

结果

在屈光手术后第1周、第4周和第12周,Orbscan II测量的平均RCCT分别为413±72、435±65和440±69微米。Pentacam测量的分别为434±51、436±53和438±50微米。术后第1周和第4周Orbscan II的测量值显著小于理论RCCT(P<0.01,P<0.01,配对t检验)。在任何阶段,理论RCCT与Pentacam测量值之间均无统计学差异。

结论

术后角膜的Orbscan II测量值往往比理论值薄,但Pentacam的测量值并非如此。