Jennings I, Luddington R J, Baglin T
Department of Haematology, Addenbrooke's Hospital, Cambridge.
J Clin Pathol. 1991 Nov;44(11):950-3. doi: 10.1136/jcp.44.11.950.
The Ciba Corning Biotrack 512 coagulation monitor requires a minimal degree of technical expertise to operate, and is already in use for near-patient testing. This study evaluated the monitor for possible use in decentralised control of oral anticoagulant treatment. The monitor compared well with Manchester Reagent, suggesting that it could be used in areas where this thromboplastin is used for centralised control. The inability of the monitor to allow for locally determined geometric mean normal prothrombin times in the calculation of the International Normalised Ratio (INR), and possibly the high International Sensitivity Index (ISI) of the thromboplastin used with the monitor, resulted in poor comparability with some other thromboplastins, particularly Thrombotest. These problems need to be addressed if the monitor is to be used for decentralised anticoagulant control.
西巴·康宁Biotrack 512凝血监测仪操作所需的技术专业程度极低,并且已用于床边检测。本研究评估了该监测仪在口服抗凝治疗分散化管理中使用的可能性。该监测仪与曼彻斯特试剂比较结果良好,表明它可用于在集中管理中使用这种凝血活酶的地区。该监测仪在计算国际标准化比值(INR)时无法考虑当地确定的几何平均正常凝血酶原时间,并且可能与该监测仪一起使用的凝血活酶的国际敏感指数(ISI)较高,导致与其他一些凝血活酶(特别是血栓试验)的可比性较差。如果要将该监测仪用于分散化抗凝管理,这些问题需要得到解决。