一起与使用受污染的氧气输送装置相关的全国性小儿患者嗜麦芽窄食单胞菌暴发。

A national outbreak of Ralstonia mannitolilytica associated with use of a contaminated oxygen-delivery device among pediatric patients.

作者信息

Jhung Michael A, Sunenshine Rebecca H, Noble-Wang Judith, Coffin Susan E, St John Keith, Lewis Felicia M, Jensen Bette, Peterson Alicia, LiPuma John, Arduino Matthew J, Holzmann-Pazgal Galit, Atkins Jane T, Srinivasan Arjun

机构信息

Division of Healthcare Quality Promotion, Centers for Disease Control and Prevention, 1600 Clifton Rd, NE MS A-24, Atlanta, GA 30333, USA.

出版信息

Pediatrics. 2007 Jun;119(6):1061-8. doi: 10.1542/peds.2006-3739.

Abstract

OBJECTIVES

In August 2005, the Centers for Disease Control and Prevention was notified of a Ralstonia species outbreak among pediatric patients receiving supplemental oxygen therapy with the Vapotherm 2000i (Vapotherm, Inc, Stevensville, MD). The Vapotherm 2000i is a reusable medical device that was used in >900 hospitals in the United States in 2005. Ralstonia are waterborne bacilli that have been implicated in hospital-acquired infections. We initiated an investigation to determine the source of the outbreak and implement infection control and prevention measures.

PATIENTS AND METHODS

We performed a case-control study at 1 hospital and conducted national case findings to obtain clinical and environmental samples for laboratory analysis. Case-patients had health care-acquired Ralstonia colonization or infection. Isolates were compared by using pulsed-field gel electrophoresis. We tested manufacturer-recommended disinfection protocols for the Vapotherm 2000i under simulated-use conditions.

RESULTS

Case-patients at the hospital (n = 5) were more likely to have received Vapotherm therapy than controls. Nationally, Ralstonia mannitolilytica was confirmed in 38 patients (aged 5 days to 7 years); 35 (92%) of the patients were exposed to the Vapotherm 2000i before recovery of the organism. Pulsed-field gel electrophoresis showed related R. mannitolilytica strains from isolates sent from 18 hospitals in 12 states. A Vapotherm machine reprocessed with a protocol proposed by the manufacturer grew Ralstonia spp after 7 days of simulated use. In December 2005, Vapotherm recalled the 2000i.

CONCLUSIONS

Our findings suggest intrinsic contamination of Vapotherm devices with Ralstonia spp. New medical devices may provide therapy equivalent to current devices yet pose novel reprocessing challenges.

摘要

目的

2005年8月,美国疾病控制与预防中心接到通知,在使用Vapotherm 2000i(Vapotherm公司,马里兰州史蒂文斯维尔)进行补充氧气治疗的儿科患者中发生了罗尔斯通氏菌属暴发。Vapotherm 2000i是一种可重复使用的医疗设备,2005年在美国900多家医院使用。罗尔斯通氏菌是水源性杆菌,与医院获得性感染有关。我们开展了一项调查,以确定暴发源头并实施感染控制与预防措施。

患者与方法

我们在1家医院进行了病例对照研究,并在全国范围内进行病例排查,以获取临床和环境样本进行实验室分析。病例患者发生了医疗保健相关的罗尔斯通氏菌定植或感染。使用脉冲场凝胶电泳对分离株进行比较。我们在模拟使用条件下测试了Vapotherm 2000i的制造商推荐消毒方案。

结果

该医院的病例患者(n = 5)比对照患者更有可能接受过Vapotherm治疗。在全国范围内,38例患者(年龄5天至7岁)确诊感染嗜麦芽窄食单胞菌;35例(92%)患者在该菌检出前接触过Vapotherm 2000i。脉冲场凝胶电泳显示,来自12个州18家医院送检的分离株中,嗜麦芽窄食单胞菌菌株具有相关性。一台按照制造商建议方案重新处理过的Vapotherm机器在模拟使用7天后培养出了罗尔斯通氏菌属。2005年12月,Vapotherm召回了2000i。

结论

我们的研究结果表明Vapotherm设备存在罗尔斯通氏菌属的内在污染。新型医疗设备可能提供与现有设备等效的治疗,但会带来新的再处理挑战。