Diffusion-weighted MRI in prostate cancer -- comparison between single-shot fast spin echo and echo planar imaging sequences.

作者信息

Kozlowski Piotr, Chang Silvia D, Goldenberg S Larry

机构信息

Prostate Centre at Vancouver General Hospital, Vancouver, British Columbia, Canada V6H 3Z6.

出版信息

Magn Reson Imaging. 2008 Jan;26(1):72-6. doi: 10.1016/j.mri.2007.04.008. Epub 2007 Jun 13.

Abstract

Diffusion-weighted (DW) MRI at 1.5 T was carried out in two groups of patients. MRI data were correlated with the biopsy and histopathology (where available). The performance of two sequences -- a single-shot FSE (14 patients) and a single-shot EPI (15 patients) -- was compared. Average ADC values from the normal peripheral zone (PZ), central gland (CG) and the tumour [prostate carcinoma (PCa)] were calculated from b values of 0 and 600. Tukey-Kramer test was used for statistical analysis. EPI produced higher values of ADC (10(-3) mm(2)/s) than FSE sequence: 1.992+/-0.208 vs. 1.573+/-0.270 in PZ (P<.001), 1.518+/-0.126 vs. 1.373+/-0.179 in CG and 1.214+/-0.254 vs. 0.993+/-0.158 in PCa (P<.01). In conclusion, both EPI and FSE sequences showed differences in ADC between normal PZ, CG and PCa; however, EPI produced significantly higher ADC values than FSE.

摘要

对两组患者进行了1.5T的扩散加权(DW)磁共振成像检查。磁共振成像数据与活检及组织病理学结果(若有)进行了关联。比较了两种序列——单次激发快速自旋回波序列(14例患者)和单次激发回波平面成像序列(15例患者)的性能。根据b值为0和600时的数据计算正常外周带(PZ)、中央腺体(CG)及肿瘤[前列腺癌(PCa)]的平均表观扩散系数(ADC)值。采用Tukey-Kramer检验进行统计学分析。回波平面成像序列产生的ADC值(10⁻³mm²/s)高于快速自旋回波序列:外周带中分别为1.992±0.208和1.573±0.270(P<0.001),中央腺体中分别为1.518±0.126和1.373±0.179,前列腺癌中分别为1.214±0.254和0.993±0.158(P<0.01)。总之,回波平面成像序列和快速自旋回波序列均显示正常外周带、中央腺体及前列腺癌之间的ADC存在差异;然而,回波平面成像序列产生的ADC值显著高于快速自旋回波序列。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验