Suppr超能文献

计算机治疗计划系统质量保证计划的实施。

Implementation of a quality assurance program for computerized treatment planning systems.

机构信息

Centro de Metrologia das Radiações, Instituto de Pesquisas Energéticas e Nucleares (IPEN/CNEN), Av. Prof Lineu Prestes 2242, CEP 05508-900, São Paulo, Brazil.

出版信息

Med Phys. 2007 Jul;34(7):2827-36. doi: 10.1118/1.2742779.

Abstract

In the present investigation, the necessary tests for implementing a quality assurance program for a commercial treatment planning system (TPS), recently installed at Sao Paulo University School of Medicine Clinicas Hospital-Brazil, was established and performed in accordance with the new IAEA publication TRS 430, and with AAPM Task Group 53. The tests recommended by those documents are classified mainly into acceptance, commissioning (dosimetric and nondosimetric), periodic quality assurance, and patient specific quality assurance tests. The recommendations of both IAEA and AAPM documents are being implemented at the hospital for photon beams produced by two linear accelerators. A Farmer ionization chamber was used in a 30 x 30 x 30 cm3 phantom with a dose rate of 320 monitor unit (MU)/min and 50 MU in the case of the dosimetric tests. The acceptance tests verified hardware, network systems integration, data transfer, and software parameters. The results obtained are in good agreement with the specifications of the manufacturer. For the commissioning dosimetric tests, the absolute dose was measured for simple geometries, such as square and rectangular fields, up to more complex geometries such as off-axis hard wedges and for behavior in the build up region. Results were analysed by the use of confidence limit as proposed by Venselaar et al. [Radio Ther. Oncol. 60, 191-201 (2001)]. Criteria of acceptability had been applied also for the comparison between the values of MU calculated manually and MU generated by TPS. The results of the dosimetric tests show that work can be reduced by choosing to perform only those that are more crucial, such as oblique incidence, shaped fields, hard wedges, and buildup region behavior. Staff experience with the implementation of the quality assurance program for a commercial TPS is extremely useful as part of a training program.

摘要

在本研究中,根据新的国际原子能机构出版物 TRS 430 和 AAPM 工作组 53,为巴西圣保罗大学医学院临床医院最近安装的商业治疗计划系统(TPS)制定并执行了实施质量保证计划的必要测试。这些文件推荐的测试主要分为验收、调试(剂量学和非剂量学)、定期质量保证和患者特定质量保证测试。国际原子能机构和 AAPM 文件的建议正在医院为两台线性加速器产生的光子束实施。在一个 30x30x30cm3 的体模中,使用 Farmer 电离室,剂量率为 320MU/min,在剂量学测试中为 50MU。验收测试验证了硬件、网络系统集成、数据传输和软件参数。获得的结果与制造商的规格非常吻合。对于调试剂量学测试,绝对剂量是在简单的几何形状(如方形和矩形场)以及更复杂的几何形状(如离轴硬楔和建成区行为)中进行测量的。结果通过 Venselaar 等人提出的置信限进行分析。[Radiother Oncol. 60, 191-201 (2001)]。对于手动计算的 MU 值和 TPS 生成的 MU 值之间的比较,也应用了可接受性标准。剂量学测试的结果表明,通过选择执行更关键的测试,可以减少工作量,例如斜入射、形状场、硬楔和建成区行为。工作人员在商业 TPS 质量保证计划实施方面的经验对于培训计划非常有用。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验