Multifocal blue-on-yellow visual evoked potentials in early glaucoma.

作者信息

Klistorner Alexander, Graham Stuart L, Martins Alessandra, Grigg John R, Arvind Hemamalini, Kumar Rajesh S, James Andrew C, Billson Francis A

机构信息

Department of Ophthalmology, Save Sight Institute, University of Sydney, Sydney, Australia.

出版信息

Ophthalmology. 2007 Sep;114(9):1613-21. doi: 10.1016/j.ophtha.2006.11.037.

Abstract

OBJECTIVE

To determine the sensitivity and specificity of blue-on-yellow multifocal visual evoked potentials (mfVEPs) in early glaucoma.

DESIGN

Cross-sectional study.

PARTICIPANTS

Fifty patients with a confirmed diagnosis of early glaucoma and 60 normal participants.

METHODS

Black-and-white mfVEPs and blue-on-yellow mfVEPs were recorded using the Accumap version 2.0 (ObjectiVision Pty. Ltd., Sydney, Australia). All patients also underwent achromatic standard automated perimetry (SAP).

MAIN OUTCOME MEASURES

Multifocal VEP amplitude and latency values in glaucoma patients were analyzed and compared with those of the normal controls.

RESULTS

Based on the definition of visual field defect, in the group of glaucomatous eyes with SAP defects, amplitude of blue-on-yellow mfVEP was abnormal in all 64 cases (100% sensitivity), whereas black-and-white mfVEP missed 5 cases (92.2% sensitivity). Generally, larger scotomata were noted on blue-on-yellow mfVEP compared with black-and-white mfVEP for the same eyes. There was high topographic correspondence between SAP and amplitude of blue-on-yellow mfVEP and significant (P<0.0001) correlation between them (correlation coefficient, 0.73). Abnormal amplitude was detected in 3 of 60 eyes of control subjects (95% specificity). There was, however, no correlation between visual field defect and latency delay in glaucoma patients. Although there was a significant difference between averaged latency of control and glaucoma eyes, values considerably overlapped.

CONCLUSIONS

The blue-on-yellow mfVEP is a sensitive and specific tool for detecting early glaucoma based on amplitude analysis.

摘要

目的

确定蓝黄双色多焦视觉诱发电位(mfVEP)在早期青光眼诊断中的灵敏度和特异性。

设计

横断面研究。

参与者

50例确诊为早期青光眼的患者和60名正常参与者。

方法

使用Accumap 2.0版本(澳大利亚悉尼ObjectiVision私人有限公司)记录黑白mfVEP和蓝黄双色mfVEP。所有患者均接受了消色差标准自动视野计检查(SAP)。

主要观察指标

分析青光眼患者的多焦VEP振幅和潜伏期值,并与正常对照组进行比较。

结果

根据视野缺损的定义,在伴有SAP缺损的青光眼组中,64例患者的蓝黄双色mfVEP振幅均异常(灵敏度100%),而黑白mfVEP漏诊5例(灵敏度92.2%)。一般来说,同一眼的蓝黄双色mfVEP上的暗点比黑白mfVEP上的更大。SAP与蓝黄双色mfVEP振幅之间存在高度的地形对应关系,且二者之间存在显著相关性(P<0.0001)(相关系数为0.73)。在60名对照者的60只眼中,有3只眼检测到异常振幅(特异性95%)。然而,青光眼患者的视野缺损与潜伏期延迟之间无相关性。尽管对照组和青光眼组的平均潜伏期存在显著差异,但数值有相当大的重叠。

结论

基于振幅分析,蓝黄双色mfVEP是检测早期青光眼的一种敏感且特异的工具。