Ozsezer Ebru, Inan Ugur, Aydin Ugur
Faculty of Dentistry, Department of Endodontics and Restorative Dentistry, Ondokuz Mayis University, Kurupelit/Samsun, Turkey.
J Endod. 2007 Aug;33(8):974-7. doi: 10.1016/j.joen.2007.04.003. Epub 2007 Jun 22.
The aim of this in vivo study was to evaluate the performance of ProPex apex locator after extirpation and in presence of different irrigation solutions: 2.5% NaOCl, 0.9% NaCl, and 0.2% chlorhexidine gluconate solutions. Twenty-seven single-rooted human teeth scheduled for extraction with mature apices were selected for this study. Measurements were performed by using ProPex after extirpation and in presence of the irrigation solutions. After extraction of the teeth, a stereomicroscope was used to confirm visually the relationship of the tip of the endodontic file to the apical foramen, and actual lengths (ALs) were determined by reducing 0.5 mm from these lengths. The data were statistically analyzed with cluster analysis. The results of this study showed that measurements after extirpation had the smallest distance to the ALs. Among the irrigation solution groups, chlorhexidine gluconate group had the smallest distance to the ALs, whereas saline group had the greatest. Most accurate measurements were obtained after extirpation.
本体内研究的目的是评估ProPex根尖定位仪在根管清理后以及存在不同冲洗液(2.5%次氯酸钠、0.9%氯化钠和0.2%葡萄糖酸氯己定溶液)的情况下的性能。本研究选取了27颗计划拔除的单根成熟根尖恒牙。在根管清理后以及存在冲洗液的情况下使用ProPex进行测量。牙齿拔除后,使用体视显微镜直观确认根管锉尖端与根尖孔的关系,并通过从这些长度中减去0.5mm来确定实际长度(AL)。数据采用聚类分析进行统计学分析。本研究结果表明,根管清理后的测量值与实际长度的距离最小。在冲洗液组中,葡萄糖酸氯己定组与实际长度的距离最小,而盐水组最大。根管清理后获得的测量最为准确。