The influence of two forms of chlorhexidine on the accuracy of contemporary electronic apex locators.

机构信息

Department of Preclinical Conservative Dentistry and Preclinical Endodontics, Pomeranian Medical University, Al. Powstańców Wielkopolskich 72, 70-111, Szczecin, Poland.

Department of Conservative Dentistry, Pomeranian Medical University, Al. Powstańców Wielkopolskich 72, 70-111, Szczecin, Poland.

出版信息

BMC Oral Health. 2019 Dec 31;20(1):3. doi: 10.1186/s12903-019-0994-z.

Abstract

BACKGROUND

Accurate determination of working length (WL) is crucial for the success of endodontic therapy. The aim of this study was to determine the influence of 2% chlorhexidine digluconate solution, 2% chlorhexidine digluconate gel and 2% hypochlorite solution on the accuracy of two devices: the Raypex 5 and the ApexDal.

METHODS

Twenty-nine single-rooted human teeth were used in this study. The crowns were cut horizontally and embedded in an alginate mass. In each tooth, six endodontic measurements were made using two electronic apex locators (EALs): a Raypex 5 and an ApexDal. For each EAL, measurements were taken with the following products: 2% chlorhexidine solution (CHX-S group), 2% chlorhexidine gel (CHX-G group) and 2% NaOCl (NaOCl group). After performing an endodontic measurement, the endodontic instruments were stabilized with flow resin composite. Afterwards, the roots were removed from the alginate mass, and the apical one-third of each root was cut lengthways to recover the canal system. Last, the distance between the file tip and the apical foramen was measured under a microscope at 60 x magnification.

RESULTS

Statistically significant differences were found between CHX-S and NaOCl and CHX-G and NaOCl, but no significant differences were detected between CHX-S and CHX-G during the testing of both devices. No statistically significant differences were observed between the Raypex 5 and ApexDal for all intracanal media tested.

CONCLUSION

The EALs Raypex 5 and ApexDal had higher accuracy in the anatomical foramen of the root containing chlorhexidine in the gel or in the solution form than in the canal containing sodium hypochlorite.

摘要

背景

准确确定工作长度(WL)对于根管治疗的成功至关重要。本研究旨在确定 2%葡萄糖酸氯己定溶液、2%葡萄糖酸氯己定凝胶和 2%次氯酸钠溶液对两种设备(Raypex 5 和 ApexDal)准确性的影响。

方法

本研究使用 29 颗单根人牙。牙冠被水平切割并嵌入藻酸盐块中。在每颗牙齿中,使用两种电子根尖定位仪(EAL)进行 6 次根管测量:Raypex 5 和 ApexDal。对于每个 EAL,使用以下产品进行测量:2%葡萄糖酸氯己定溶液(CHX-S 组)、2%葡萄糖酸氯己定凝胶(CHX-G 组)和 2%次氯酸钠(NaOCl 组)。进行根管测量后,用流动树脂复合材料固定根管器械。然后,将根从藻酸盐块中取出,将每个根的根尖三分之一纵向切割以恢复根管系统。最后,在 60x 放大倍数的显微镜下测量文件尖端和根尖孔之间的距离。

结果

在测试两种设备时,CHX-S 与 NaOCl 之间以及 CHX-G 与 NaOCl 之间存在统计学上的显著差异,但 CHX-S 与 CHX-G 之间未检测到显著差异。在测试的所有根管内介质中,Raypex 5 和 ApexDal 之间均未观察到统计学上的显著差异。

结论

EALs Raypex 5 和 ApexDal 在含有凝胶或溶液形式氯己定的根的解剖学根尖孔中的准确性高于含有次氯酸钠的根管。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/8c3d/6938632/08486b1fce3c/12903_2019_994_Fig1_HTML.jpg

相似文献

1
The influence of two forms of chlorhexidine on the accuracy of contemporary electronic apex locators.
BMC Oral Health. 2019 Dec 31;20(1):3. doi: 10.1186/s12903-019-0994-z.
2
In vivo evaluation of the iPex and Root ZX electronic apex locators using various irrigants.
Int Endod J. 2013 Aug;46(8):769-74. doi: 10.1111/iej.12057. Epub 2013 Mar 28.
3
In vivo evaluation of the Raypex 5 by using different irrigants.
J Endod. 2012 Aug;38(8):1075-7. doi: 10.1016/j.joen.2012.05.018.
4
An ex vivo comparison of working length determination by 3 electronic apex locators.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009 Sep;108(3):e147-51. doi: 10.1016/j.tripleo.2009.04.040.
5
A comparison between two electronic apex locators: an ex vivo investigation.
Int Endod J. 2007 May;40(5):362-73. doi: 10.1111/j.1365-2591.2006.01229.x. Epub 2007 Feb 19.
6
Accuracy of a new apex locator: an in vitro study.
Int Endod J. 2002 Feb;35(2):186-92. doi: 10.1046/j.1365-2591.2002.00468.x.
8
In vivo accuracy of three electronic root canal length measurement devices: Dentaport ZX, Raypex 5 and ProPex II.
Int Endod J. 2012 Jun;45(6):552-6. doi: 10.1111/j.1365-2591.2011.02010.x. Epub 2012 Jan 19.
9
Accuracy of electronic apex locator in enlarged root canals with different root canal irrigants: an in vitro study.
J Contemp Dent Pract. 2013 Jul 1;14(4):649-52. doi: 10.5005/jp-journals-10024-1379.
10
Evaluation of precision of length determination with 3 electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator, and RomiAPEX D-30.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007 Oct;104(4):e91-4. doi: 10.1016/j.tripleo.2007.05.016. Epub 2007 Aug 20.

引用本文的文献

1
Factors influencing the accuracy of electronic apex locators: A scoping review.
Dent Res J (Isfahan). 2025 Jun 27;22:26. doi: 10.4103/drj.drj_634_23. eCollection 2025.
2
The effect of different root canal irrigations on the accuracy of apex locators: A systematic review.
J Clin Exp Dent. 2024 Dec 1;16(12):e1538-e1546. doi: 10.4317/jced.62049. eCollection 2024 Dec.
3
Effect of Different Electroconductive Root Canal Irrigations on the Accuracy of Different Apex Locators: An Comparative Study.
J Int Soc Prev Community Dent. 2024 Jun 27;14(3):211-217. doi: 10.4103/jispcd.jispcd_192_23. eCollection 2024 May-Jun.
4
The Effect of Coronal Pre-flaring and Type of Root Canal Irrigation on Working Length Accuracy Using Electronic Apex Locators.
F1000Res. 2023 Oct 24;12:533. doi: 10.12688/f1000research.133288.3. eCollection 2023.
6
The Effect of Long Service Life on the Accuracy of Dentaport ZX Electronic Apex Locator.
Iran Endod J. 2022;17(4):195-199. doi: 10.22037/iej.v17i4.34623.
7
Efficiency of an Integrated Apex Locator in Determining Working Length in Various Irrigating Solutions: An Study.
J Pharm Bioallied Sci. 2020 Aug;12(Suppl 1):S410-S414. doi: 10.4103/jpbs.JPBS_122_20. Epub 2020 Aug 28.

本文引用的文献

2
A comparative assessment of the accuracies of Raypex 5, Raypex 6, iPex and iPex II electronic apex locators: An in vitro study.
J Istanb Univ Fac Dent. 2017 Jan 2;51(1):28-33. doi: 10.17096/jiufd.61309. eCollection 2017.
3
Determination of the Accuracy of 5 Electronic Apex Locators in the Function of Different Employment Protocols.
J Endod. 2017 Oct;43(10):1663-1667. doi: 10.1016/j.joen.2017.03.039. Epub 2017 Jul 8.
6
Influence of Blood on the Accuracy of Raypex 5 and Root ZX Electronic Foramen Locators: An In Vivo Study.
Braz Dent J. 2016 May-Jun;27(3):336-9. doi: 10.1590/0103-6440201600678.
9
Comparative evaluation of accuracy of 2 electronic Apex locators with conventional radiography: an ex vivo study.
J Endod. 2015 Feb;41(2):201-4. doi: 10.1016/j.joen.2014.10.011. Epub 2014 Dec 2.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验