Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex.

作者信息

Plotino G, Grande N M, Brigante L, Lesti B, Somma F

机构信息

Department of Endodontics, School of Dentistry, Catholic University of Sacred Heart, Rome, Italy.

出版信息

Int Endod J. 2006 May;39(5):408-14. doi: 10.1111/j.1365-2591.2006.01095.x.

Abstract

AIM

To compare ex vivo the accuracy of three electronic apex locators (EALs): Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex.

METHODOLOGY

Electronic working length determination was carried out in 40 extracted teeth using an ex vivo model. After access preparation, a first operator determined the reference length (AL) for each tooth under a 30x stereomicroscope using the apical constriction as the apical landmark. All teeth were then measured with each EAL and the results obtained were compared with the corresponding AL. The AL was subtracted from the electronically determined distance. The measurements exceeding the AL were recorded as positive (long) and the measurements short of the AL were recorded as negative. Data were analyzed using the Friedman Test and Tukey multiple range test for nonparametric correlation amongst groups. Statistical significance was considered at P < 0.05.

RESULTS

Comparing the differences between measurements obtained with the three EALs and those obtained with the stereomicroscope, the percentage of measurements within +/-0.5 mm of the AL was 97.37% (84.22% within 0.5 mm short of AL) for the Root ZX, 94.28% (88.57% within 0.5 mm short of AL) for the Elements and 100% (35.9% within 0.5 mm short of AL) for the ProPex. The mean difference between the AL and the lengths measured by the Root ZX, the Elements and the ProPex were, respectively, -0.157 +/- 0.228, -0.103 +/- 0.359 and 0.307 +/- 0.271 mm.

CONCLUSIONS

The results of the present study confirm that the EALs determined the canal length within +/-0.5 mm from the apical constriction in the majority of cases. The majority of the ProPex readings were long.

摘要

目的

在体外比较三种电子根尖定位仪(EAL):Root ZX、Elements诊断仪和根尖定位仪ProPex的准确性。

方法

采用体外模型,对40颗离体牙进行电子工作长度测定。开髓预备后,由一名操作人员在30倍体视显微镜下以根尖狭窄处作为根尖标志确定每颗牙的参考长度(AL)。然后用每种EAL对所有牙齿进行测量,并将所得结果与相应的AL进行比较。用电子测定距离减去AL。超过AL的测量值记为正(长),低于AL的测量值记为负。采用Friedman检验和Tukey多重极差检验对组间非参数相关性进行数据分析。P<0.05认为具有统计学意义。

结果

比较三种EAL测量结果与体视显微镜测量结果的差异,Root ZX在AL的±0.5mm范围内测量值的百分比为97.37%(低于AL 0.5mm范围内为84.22%),Elements为94.28%(低于AL 0.5mm范围内为88.57%),ProPex为100%(低于AL 0.5mm范围内为35.9%)。Root ZX、Elements和ProPex测量长度与AL的平均差值分别为-0.157±0.228、-0.103±0.359和0.307±0.271mm。

结论

本研究结果证实,在大多数情况下,EAL能在距根尖狭窄处±0.5mm范围内确定根管长度。ProPex的大多数读数偏长。