Schneider Miklos, Borgulya Gabor, Seres Andras, Nagy Zoltan, Nemeth Janos
Department of Ophthalmology, Semmelweis University, Faculty of Medicine, Budapest - Hungary.
Eur J Ophthalmol. 2009 Mar-Apr;19(2):180-7. doi: 10.1177/112067210901900202.
To compare central corneal thickness measurements made using two different methods, optical coherence tomography (OCT) and ultrasound pachymetry, applied both in normal eyes and in eyes on which photorefractive keratectomy (PRK) had been performed. A second objective was to assess the intrasession variability of OCT measurements.
In this prospective study, central corneal thickness was measured in 20 normal subjects (normal group) and in 20 PRK patients using the StratusOCT instrument model 3000 (Carl Zeiss Meditec), and also with an ultrasound pachymeter. Five OCT measurements were performed using the Fast Macular Thickness protocol. Corneal thickness data were obtained with the Scan Profile analysis protocol. The OCT measurement results were compared with the mean value of three ultrasound pachymetry measurements for the same eye.
The pachymetry-OCT correlation coefficients were 0.96 and 0.97 in the normal and PRK groups, respectively (p=0.14). Neither linear regression nor Bland-Altmann analysis revealed any significant systematic measurement error. Intrasession standard deviations in the normal and PRK groups were 4.9 microm and 3.8 microm, respectively.
Noncontact central corneal thickness measurements made using the StratusOCT instrument are accurate and reproducible, both in normal subjects and in post-PRK patients. The instrument system does not need any modifications to correctly detect and measure the center of the cornea.
比较使用光学相干断层扫描(OCT)和超声测厚法这两种不同方法测量正常眼及接受过准分子激光角膜切削术(PRK)的眼睛的中央角膜厚度。第二个目的是评估OCT测量的测量期间变异性。
在这项前瞻性研究中,使用Stratus OCT 3000型仪器(卡尔蔡司医疗技术公司)以及超声测厚仪,对20名正常受试者(正常组)和20名PRK患者测量中央角膜厚度。使用快速黄斑厚度方案进行5次OCT测量。通过扫描轮廓分析方案获取角膜厚度数据。将OCT测量结果与同一只眼睛的三次超声测厚测量的平均值进行比较。
正常组和PRK组中测厚法与OCT的相关系数分别为0.96和0.97(p = 0.14)。线性回归和布兰德-奥特曼分析均未显示任何显著的系统测量误差。正常组和PRK组的测量期间标准差分别为4.9微米和3.8微米。
使用Stratus OCT仪器进行的非接触式中央角膜厚度测量在正常受试者和PRK术后患者中都是准确且可重复的。该仪器系统无需任何修改即可正确检测和测量角膜中心。