Hruby Gregory W, Marruffo Franzo C, Durak Evren, Collins Sean M, Pierorazio Phillip, Humphrey Peter A, Mansukhani Mahesh M, Landman Jaime
Department of Urology, Columbia University Medical Center, New York, New York, USA.
J Urol. 2007 Dec;178(6):2689-93. doi: 10.1016/j.juro.2007.07.121. Epub 2007 Oct 22.
Vessel sealing capabilities and the peripheral energy spread associated with currently available energy based surgical instruments were evaluated.
Four groups of energy based surgical instruments were established for evaluation, including the ACE, LCS-C5 (Ethicon, Cincinnati, Ohio), LigaSure V and the prototype Trissector. In vivo vessel diameter under physiological conditions, bursting pressures and seal failure were recorded and analyzed for arteries and veins separately. After burst testing the vessels were fixed and evaluated histopathologically for peripheral energy damage. ANOVA was used to determine differences between groups.
The LigaSure V sealed arteries and veins with an average burst pressure of 536 and 386 mm Hg, respectively. The Harmonic ACE sealed arteries and veins with an average burst pressure of 436 and 160 mm Hg, respectively. The Harmonic LCS-C5 sealed arteries and veins with an average burst pressure of 363 and 215 mm Hg, respectively. The Trissector sealed arteries and veins with an average burst pressure of 328 and 237 mm Hg, respectively. For arteries the ACE, LCS-C5, Trissector and LigaSure V measured a full-thickness peripheral energy spread of 0.6, 0.3, 8.0 and 4.5 mm, respectively (p <0.0001). For veins the ACE, LCS-C5, Trissector and LigaSure V measured a full-thickness peripheral energy spread of 1.5, 1.3, 8.5 and 6.3 mm, respectively (p = 0.003).
The LigaSure 5 was superior to the other devices tested regarding the ability to seal vessels up to 7 mm. The ACE is an efficient vessel sealing system with the ability to seal vessels up to 5 mm.
评估当前可用的基于能量的手术器械的血管封闭能力及周边能量扩散情况。
建立四组基于能量的手术器械进行评估,包括ACE、LCS-C5(爱惜康公司,俄亥俄州辛辛那提)、LigaSure V和原型Trissector。分别记录并分析生理条件下动脉和静脉的体内血管直径、破裂压力及密封失败情况。破裂测试后,将血管固定并进行组织病理学评估以确定周边能量损伤。采用方差分析确定组间差异。
LigaSure V封闭动脉和静脉的平均破裂压力分别为536和386 mmHg。Harmonic ACE封闭动脉和静脉的平均破裂压力分别为436和160 mmHg。Harmonic LCS-C5封闭动脉和静脉的平均破裂压力分别为363和215 mmHg。Trissector封闭动脉和静脉的平均破裂压力分别为328和237 mmHg。对于动脉,ACE、LCS-C5、Trissector和LigaSure V的全层周边能量扩散分别为0.6、0.3、8.0和4.5 mm(p<0.0001)。对于静脉,ACE、LCS-C5、Trissector和LigaSure V的全层周边能量扩散分别为1.5、1.3、8.5和6.3 mm(p = 0.003)。
在封闭直径达7 mm的血管的能力方面,LigaSure 5优于其他测试器械。ACE是一种高效的血管封闭系统,能够封闭直径达5 mm的血管。