在健康受试者和胃轻瘫受试者中,不可消化胶囊与放射性标记餐的胃排空情况比较。

Comparison of gastric emptying of a nondigestible capsule to a radio-labelled meal in healthy and gastroparetic subjects.

作者信息

Kuo B, McCallum R W, Koch K L, Sitrin M D, Wo J M, Chey W D, Hasler W L, Lackner J M, Katz L A, Semler J R, Wilding G E, Parkman H P

机构信息

Gastroenterology Unit, Massachusetts General Hospital, Harvard Medical School, Boston, MA 02114, USA.

出版信息

Aliment Pharmacol Ther. 2008 Jan 15;27(2):186-96. doi: 10.1111/j.1365-2036.2007.03564.x. Epub 2007 Oct 28.

Abstract

BACKGROUND

Gastric emptying scintigraphy (GES) using a radio-labelled meal is used to measure gastric emptying. A nondigestible capsule, SmartPill, records luminal pH, temperature, and pressure during gastrointestinal transit providing a measure of gastric emptying time (GET).

AIMS

To compare gastric emptying time and GES by assessing their correlation, and to compare GET and GES for discriminating healthy subjects from gastroparetics.

METHODS

Eighty-seven healthy subjects and 61 gastroparetics enrolled with simultaneous SmartPill and GES. Fasted subjects were ingested capsule and [(99m)Tc]-SC radio-labelled meal. Images were obtained every 30 min for 6 h. Gastric emptying time and percentage of meal remaining at 2/4 h were determined for each subject. The sensitivity/specificity and receiver operating characteristic analysis of each measure were determined for each subject.

RESULTS

Correlation between GET and GES-4 h was 0.73 and GES-2 h was 0.63. The diagnostic accuracy from the receiver operating characteristic curve between gastroparetics and healthy subjects was GET = 0.83, GES-4 h = 0.82 and GES-2 h = 0.79. The 300-min cut-off time for GET gives sensitivity of 0.65 and specificity of 0.87 for diagnosis of gastroparesis. The corresponding sensitivity/specificity for 2 and 4 h standard GES measures were 0.34/0.93 and 0.44/0.93, respectively.

CONCLUSION

SmartPill GET correlates with GES and discriminates between healthy and gastroparetic subjects offering a nonradioactive, standardized, ambulatory alternative to scintigraphy.

摘要

背景

使用放射性标记餐的胃排空闪烁扫描术(GES)用于测量胃排空。一种不可消化的胶囊,智能药丸(SmartPill),在胃肠道转运过程中记录腔内pH值、温度和压力,从而提供胃排空时间(GET)的测量值。

目的

通过评估胃排空时间(GET)和GES的相关性来比较两者,并比较GET和GES在区分健康受试者和胃轻瘫患者方面的作用。

方法

87名健康受试者和61名胃轻瘫患者同时接受智能药丸和GES检查。空腹受试者摄入胶囊和[(99m)Tc]-硫胶体放射性标记餐。每30分钟获取一次图像,共6小时。确定每个受试者的胃排空时间以及2/4小时时剩余餐食的百分比。对每个受试者的每项测量指标进行敏感性/特异性和受试者工作特征分析。

结果

GET与4小时GES的相关性为0.73,与2小时GES的相关性为0.63。胃轻瘫患者和健康受试者之间受试者工作特征曲线的诊断准确性为:GET = 0.83,4小时GES = 0.82,2小时GES = 0.79。GET的300分钟截断时间诊断胃轻瘫的敏感性为0.65,特异性为0.87。2小时和4小时标准GES测量的相应敏感性/特异性分别为0.34/0.93和0.44/0.93。

结论

智能药丸GET与GES相关,并且能够区分健康受试者和胃轻瘫患者,为闪烁扫描术提供了一种非放射性、标准化的动态替代方法。