Kim Jin-Hyoung, Lee Doh, Rhee Kang-Il
Department of Ophthalmology, Ilsan Paik Hospital, Inje University College of Medicine, Goyang-si, Korea.
J Cataract Refract Surg. 2008 Jan;34(1):132-6. doi: 10.1016/j.jcrs.2007.08.036.
To compare ultrasound (US) pachymetry, Orbscan, and optical coherence tomography (OCT) measurements of the central corneal thickness (CCT) before laser in situ keratomileusis (LASIK) and evaluate the reproducibility of flaps created with the IntraLase femtosecond laser (IntraLase, Inc.) using OCT.
Department of Ophthalmology, IIsan Paik Hospital, Goyang, Korea.
Central corneal thickness was measured using OCT, US pachymetry, and Orbscan in 59 eyes of 30 patients before femtosecond laser LASIK. The postoperative corneal flap thickness, measured using OCT, was compared with the preoperative intended thickness.
Optical coherence tomography, US pachymetry, and Orbscan measurements provided similar CCT values (P>.05). Corneal thickness values obtained using US or Orbscan correlated well with those obtained by OCT, with the correlation coefficient ranging from 0.804 to 0.889 (P<.05). The OCT measurements showed no significant difference between the postoperative flap thickness and the intended flap thickness (P>.05).
Optical coherence tomography was comparable to US pachymetry and Orbscan in cornea thickness measurement. Optical coherence tomography was easy and relatively accurate to use preoperatively and in the early postoperative period. The femtosecond laser created highly reproducible flaps that corresponded with the preoperative intended thickness.
比较准分子原位角膜磨镶术(LASIK)术前超声(US)角膜测厚、Orbscan和光学相干断层扫描(OCT)测量中央角膜厚度(CCT)的结果,并使用OCT评估IntraLase飞秒激光(IntraLase公司)制作的角膜瓣的可重复性。
韩国高阳市峨山白医院眼科。
在30例患者的59只眼中,于飞秒激光LASIK术前使用OCT、US角膜测厚法和Orbscan测量中央角膜厚度。使用OCT测量术后角膜瓣厚度,并与术前预期厚度进行比较。
光学相干断层扫描、US角膜测厚法和Orbscan测量得到的CCT值相似(P>0.05)。使用US或Orbscan获得的角膜厚度值与OCT获得的值相关性良好,相关系数在0.804至0.889之间(P<0.05)。OCT测量显示术后瓣厚度与预期瓣厚度之间无显著差异(P>0.05)。
光学相干断层扫描在角膜厚度测量方面与US角膜测厚法和Orbscan相当。光学相干断层扫描在术前和术后早期使用简便且相对准确。飞秒激光制作的角膜瓣具有高度可重复性,与术前预期厚度相符。