Corneal parameters and difference between goldmann applanation tonometry and dynamic contour tonometry in normal eyes.

作者信息

Lanza Michele, Borrelli Maria, De Bernardo Maddalena, Filosa Maria Luisa, Rosa Nicola

机构信息

Eye Department, Centro Grandi Apparecchiature, Second University of Naples, Naples, Italy.

出版信息

J Glaucoma. 2008 Sep;17(6):460-4. doi: 10.1097/IJG.0b013e31816224bd.

Abstract

PURPOSE

To compare the difference between 2 methods of measuring the intraocular pressure (IOP), namely, Goldmann applanation tonometry (GAT) and dynamic contour tonometry (DCT) and to study the relationship with the corneal parameters in patients with no glaucoma signs.

METHODS

One hundred and eighteen eyes of 118 healthy subjects from 20 to 77 years of age underwent IOP measurements with DCT and GAT, central corneal thickness, and corneal volume measurements with Pentacam Scheimpflug camera. Pentacam examination has been performed first, then DCT and, after 10 minutes, GAT measurements.

RESULTS

The measurements with GAT ranged between 11 and 22 mm Hg (m=15.49+/-2.43 mm Hg), the measurements with DCT ranged between 10.5 and 25.1 mm Hg (m=17.59+/-2.9 mm Hg). DCT showed a statistically significant (P<0.0001) higher IOP measurement compared with GAT. A relation between differences in IOP with central corneal thickness and corneal volume was found, whereas no relation was found with corneal radius and age.

CONCLUSIONS

Our results show a discrete correlation between GAT and DCT measurements, but DCT showed slightly higher values of IOP. If DCT should be considered the gold standard, higher values of IOP could still be considered normal. These 2 devices cannot to be used interchangeably.

摘要

目的

比较两种测量眼压(IOP)的方法,即Goldmann压平眼压计(GAT)和动态轮廓眼压计(DCT)之间的差异,并研究无青光眼体征患者的眼压与角膜参数之间的关系。

方法

对118名年龄在20至77岁之间的健康受试者的118只眼睛进行了DCT和GAT眼压测量、中央角膜厚度测量以及使用Pentacam Scheimpflug相机进行角膜体积测量。首先进行Pentacam检查,然后进行DCT测量,10分钟后进行GAT测量。

结果

GAT测量值在11至22毫米汞柱之间(平均值=15.49±2.43毫米汞柱),DCT测量值在10.5至25.1毫米汞柱之间(平均值=17.59±2.9毫米汞柱)。与GAT相比,DCT显示出具有统计学意义(P<0.0001)的更高眼压测量值。发现眼压差异与中央角膜厚度和角膜体积之间存在关系,而与角膜半径和年龄无关。

结论

我们的结果显示GAT和DCT测量值之间存在离散相关性,但DCT显示的眼压值略高。如果将DCT视为金标准,较高的眼压值仍可被视为正常。这两种设备不能互换使用。