Suppr超能文献

根据国际协议对欧姆龙M6(HEM - 7001 - E)上臂式血压测量装置在老年患者中的验证。

Validation of the Omron M6 (HEM-7001-E) upper arm blood pressure measuring device according to the International Protocol in elderly patients.

作者信息

Altunkan Sekip, Iliman Nevzat, Altunkan Erkan

机构信息

Hypertension Division, Metropol Medical Center, Dişkapi Yildirim Beyazit Education and Research Hospital, Ankara, Turkey.

出版信息

Blood Press Monit. 2008 Apr;13(2):117-22. doi: 10.1097/MBP.0b013e3282f3fb50.

Abstract

OBJECTIVE

Despite the widespread use of automated self-measurement monitors, there is limited published evidence on their accuracy and reliability on different patient groups. The objective of this study was to evaluate the accuracy and reliability of the Omron M6 (HEM-7001-E) upper-arm blood pressure (BP) device against mercury sphygmomanometer on elderly patients according to the criteria of the International Protocol.

DESIGNS AND METHODS

Thirty-three patients above 65 years of age, who were classified based on the BP categories of the International Protocol, were recruited for the study. BP measurements at the upper arm with the Omron M6 were compared with the results obtained by two trained observers using a mercury sphygmomanometer. Nine sequential BP measurements were taken. During the validation study, 99 measurements were obtained from 33 patients for comparison. The first phase was carried out on 15 patients and if the device passed this phase, 18 more patients were selected.

RESULTS

Mean discrepancies and standard deviations of the device sphygmomanometer were 1.4+/-5.3 mmHg for systolic BP (SBP) and -1.4+/-4.5 mmHg for diastolic BP (DBP) in the study group. The device passed phase 1 in 15 patients. In phase 2.1, from the total 99 comparisons, 76, 92, and 97 for SBP and 77, 94, and 99 for DBP were less than 5, 10, and 15 mmHg, respectively. The Omron M6 passed phases 2.1 and 2.2 in the elderly group of patients.

CONCLUSION

The Omron M6 (HEM-7001-E) upper-arm BP monitor passed according to the International Protocol criteria and can be recommended for use in elderly patients.

摘要

目的

尽管自动自我测量监测仪已被广泛使用,但关于其在不同患者群体中的准确性和可靠性的已发表证据有限。本研究的目的是根据国际协议的标准,评估欧姆龙M6(HEM - 7001 - E)上臂血压(BP)设备相对于汞柱式血压计在老年患者中的准确性和可靠性。

设计与方法

招募了33名65岁以上的患者,这些患者根据国际协议的血压类别进行分类,以参与本研究。将使用欧姆龙M6在上臂进行的血压测量结果与两名经过培训的观察者使用汞柱式血压计获得的结果进行比较。进行了九次连续的血压测量。在验证研究期间,从33名患者中获得了99次测量结果用于比较。第一阶段对15名患者进行,如果该设备通过此阶段,则再选择18名患者。

结果

在研究组中,该设备与血压计收缩压(SBP)的平均差异和标准差为1.4±5.3 mmHg,舒张压(DBP)为 - 1.4±4.5 mmHg。该设备在15名患者中通过了第一阶段。在第二阶段1中,在总共99次比较中,SBP小于5、10和15 mmHg的分别有76、92和97次,DBP小于5、10和15 mmHg的分别有77、94和99次。欧姆龙M6在老年患者组中通过了第二阶段1和第二阶段2。

结论

欧姆龙M6(HEM - 7001 - E)上臂血压监测仪根据国际协议标准通过了测试,可推荐用于老年患者。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验