Capillary and venous blood glucose measurements using a direct glucose-sensing meter.

作者信息

Matthews D R, Burton S F, Bown E, Chusney G, Dornan T, Gale E A, McKinnon G, Steemson J

机构信息

Diabetes Research Laboratories, Radcliffe Infirmary, Oxford, UK.

出版信息

Diabet Med. 1991 Nov;8(9):875-80. doi: 10.1111/j.1464-5491.1991.tb02128.x.

Abstract

The aim of the study was to evaluate the precision and accuracy of the ExacTech home blood glucose meter when used with either capillary or venous blood and to compare this with a reference whole blood glucose assay. Non-fasting glucose measurements were used since a validation study showed no capillary-venous differences between fasting and post-prandial states. In a cross-sectional study, blood was taken from 182 patients and measured in duplicate on three batches of strips. Altogether we analysed 1089 readings. The regression of the data from capillary blood samples (meter vs reference method) had a correlation coefficient, of 0.93, and a mean bias of 0.2 mmol l-1. The corrected 90% confidence interval was +/- 1.5 mmol l-1 overall, and +/- 0.9 mmol l-1 for readings under 7.0 mmol l-1. Regression of the data from venous blood samples (meter vs reference method) had a correlation coefficient of 0.93 and a slope of x 1.1. The corrected 90% confidence interval was +/- 1.7 mmol l-1. Thus venous blood may be used even though the meter is calibrated for capillary samples but the value must be corrected by dividing by 1.1. Error-grid analysis showed that day-to-day clinical decisions could be made on the basis of ExacTech readings, although a diagnosis of borderline diabetes may not be possible.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验