Spectral analysis of noise in the neonatal intensive care unit.

作者信息

Livera M D, Priya B, Ramesh A, Suman Rao P N, Srilakshmi V, Nagapoornima M, Ramakrishnan A G, Dominic M

机构信息

St John's Medical College Hospital, Bangalore, India.

出版信息

Indian J Pediatr. 2008 Mar;75(3):217-22. doi: 10.1007/s12098-008-0048-z.

Abstract

OBJECTIVE

To perform spectral analysis of noise generated by equipments and activities in a level III neonatal intensive care unit (NICU) and measure the real time sequential hourly noise levels over a 15 day period.

METHODS

Noise generated in the NICU by individual equipments and activities were recorded with a digital spectral sound analyzer to perform spectral analysis over 0.5 - 8 KHz. Sequential hourly noise level measurements in all the rooms of the NICU were done for 15 days using a digital sound pressure level meter . Independent sample t test and one way ANOVA were used to examine the statistical significance of the results. The study has a 90 % power to detect at least 4 dB differences from the recommended maximum of 50 dB with 95 % confidence.

RESULTS

The mean noise levels in the ventilator room and stable room were 19.99 dB (A) sound pressure level (SPL) and 11.81 dB (A) SPL higher than the maximum recommended of 50 dB (A) respectively ( p < 0.001). The equipments generated 19.11 dB SPL higher than the recommended norms in 1 - 8 KHz spectrum. The activities generated 21.49 dB SPL higher than the recommended norms in 1 - 8 KHz spectrum ( p< 0.001). The ventilator and nebulisers produced excess noise of 8.5 dB SPL at the 0.5 KHz spectrum.

CONCLUSION

Noise level in the NICU is unacceptably high .Spectral analysis of equipment and activity noise have shown noise predominantly in the 1 - 8 KHz spectrum. These levels warrant immediate implementation of noise reduction protocols as a standard of care in the NICU.

摘要

目的

对三级新生儿重症监护病房(NICU)中设备及活动产生的噪声进行频谱分析,并在15天内测量实时逐小时噪声水平。

方法

使用数字频谱声音分析仪记录NICU中单个设备及活动产生的噪声,以在0.5 - 8千赫兹范围内进行频谱分析。使用数字声压级计对NICU所有房间进行连续15天的逐小时噪声水平测量。采用独立样本t检验和单因素方差分析来检验结果的统计学意义。该研究有90%的把握度在95%置信度下检测出与推荐最大值50分贝至少相差4分贝的差异。

结果

通风机房和稳定病房的平均噪声水平分别比推荐最大值50分贝(A)的声压级(SPL)高19.99分贝(A)和11.81分贝(A)(p < 0.001)。设备在1 - 8千赫兹频谱中产生的声压级比推荐标准高19.11分贝。活动在1 - 8千赫兹频谱中产生的声压级比推荐标准高21.49分贝(p < 0.001)。通风机和雾化器在0.5千赫兹频谱处产生了8.5分贝声压级的过量噪声。

结论

NICU中的噪声水平高得令人无法接受。设备及活动噪声的频谱分析表明,噪声主要集中在1 - 8千赫兹频谱。这些水平表明应立即实施降噪方案,作为NICU的护理标准。