Suppr超能文献

基于手指动脉压重建肱动脉压的验证

Validation of brachial artery pressure reconstruction from finger arterial pressure.

作者信息

Guelen Ilja, Westerhof Berend E, van der Sar Gertrude L, van Montfrans Gert A, Kiemeneij Ferdinand, Wesseling Karel H, Bos Willem Jan W

机构信息

BMEYE, Academic Medical Center, Amsterdam, The Netherlands.

出版信息

J Hypertens. 2008 Jul;26(7):1321-7. doi: 10.1097/HJH.0b013e3282fe1d28.

Abstract

OBJECTIVE

Measurement of finger artery pressure with Finapres offers noninvasive continuous blood pressure, which, however, differs from brachial artery pressure. Generalized waveform filtering and level correction may convert the finger artery pressure waveform to a brachial waveform. An upper-arm cuff return-to-flow measurement may be used to calibrate the blood pressure on an individual basis. We tested these corrective methods as implemented in the Finometer device.

METHODS

Intrabrachial artery pressure (BAP) and finger artery pressures were recorded simultaneously in 37 cardiac patients, aged 41-83 years, who underwent a cardiac catheterization procedure. Finger artery pressures were compared after waveform filtering and level correction and after an additional return-to-flow calibration. Measurements were performed in supine and sitting positions. Accuracy and precision were considered clinically acceptable if the mean and standard deviation of the return-to-flow intrabrachial artery pressure (reBAP)-BAP differences were smaller than 5 +/- 8 mmHg (Association for the Advancement of Medical Instrumentation requirements).

RESULTS

Finger artery systolic, diastolic and mean pressures for the group differed from that of intrabrachial artery pressure by -10 +/- 13, -12 +/- 8 and -16 +/- 8 mmHg, respectively. After waveform filtering and level correction the filtered level corrected arterial pressure differed by -1 +/- 11, -0 +/- 7 and -2 +/- 7 mmHg. After individual calibration, reBAP differed by 3 +/- 8, 4 +/- 6 and 3 +/- 5 mmHg. Comparable results were found in the sitting position but only when the supine return-to-flow calibration was used.

CONCLUSION

Reconstruction of intrabrachial artery pressure from finger artery pressure with waveform filtering and level correction reduces the pressure differences substantially, with diastolic and mean within Association for the Advancement of Medical Instrumentation requirements. After one supine return-to-flow calibration, all pressure differences meet the requirements. Return-to-flow calibration should not be repeated in sitting position.

摘要

目的

使用Finapres测量手指动脉压可提供无创连续血压,但该血压与肱动脉压不同。广义波形滤波和水平校正可将手指动脉压波形转换为肱动脉波形。上臂袖带回流血流测量可用于个体血压校准。我们测试了Finometer设备中实施的这些校正方法。

方法

对37例年龄在41 - 83岁、接受心脏导管插入术的心脏病患者同时记录肱动脉内压(BAP)和手指动脉压。在进行波形滤波和水平校正后以及额外的回流血流校准后,比较手指动脉压。测量在仰卧位和坐位进行。如果回流血流肱动脉内压(reBAP)与BAP差值的平均值和标准差小于5±8 mmHg(医疗器械促进协会要求),则认为准确性和精密度在临床上可接受。

结果

该组手指动脉收缩压、舒张压和平均压与肱动脉内压的差值分别为 -10±13、-12±8和 -16±8 mmHg。经过波形滤波和水平校正后,滤波后的水平校正动脉压差值为 -1±11、-0±7和 -2±7 mmHg。个体校准后,reBAP差值为3±8、4±6和3±5 mmHg。在坐位也发现了类似结果,但仅当使用仰卧位回流血流校准。

结论

通过波形滤波和水平校正从手指动脉压重建肱动脉内压可大幅降低压力差值,舒张压和平均压符合医疗器械促进协会要求。一次仰卧位回流血流校准后,所有压力差值均符合要求。坐位不应重复回流血流校准。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验