Suppr超能文献

用于质量保证应用的塑料水等效闪烁光纤剂量计阵列的临床原型。

Clinical prototype of a plastic water-equivalent scintillating fiber dosimeter array for QA applications.

作者信息

Lacroix Fréderic, Archambault Louis, Gingras Luc, Guillot Mathieu, Beddar A Sam, Beaulieu Luc

机构信息

Département de Physique, Université Laval, Québec, Québec, Canada.

出版信息

Med Phys. 2008 Aug;35(8):3682-90. doi: 10.1118/1.2953564.

Abstract

A clinical prototype of a scintillating fiber dosimeter array for quality assurance applications is presented. The array consists of a linear array of 29 plastic scintillation detectors embedded in a water-equivalent plastic sheet coupled to optical fibers used to guide optical photons to a charge coupled device (CCD) camera. The CCD is packaged in a light-tight, radiation-shielded housing designed for convenient transport. A custom designed connector is used to ensure reproducible mechanical positioning of the optical fibers relative to the CCD. Profile and depth dose characterization measurements are presented and show that the prototype provides excellent dose measurement reproducibility (+/-0.8%) in-field and good accuracy (+/-1.6% maximum deviation) relative to the dose measured with an IC10 ionization chamber.

摘要

本文介绍了一种用于质量保证应用的闪烁光纤剂量计阵列的临床原型。该阵列由29个塑料闪烁探测器的线性阵列组成,这些探测器嵌入到与光纤耦合的水等效塑料片中,光纤用于将光光子引导至电荷耦合器件(CCD)相机。CCD封装在一个设计用于方便运输的不透光、辐射屏蔽外壳中。使用定制设计的连接器来确保光纤相对于CCD的机械定位可重复。给出了剖面和深度剂量特性测量结果,结果表明该原型在射野内提供了出色的剂量测量重复性(±0.8%),并且相对于用IC10电离室测量的剂量具有良好的准确性(最大偏差±1.6%)。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验