Suppr超能文献

动力型隐形眼镜透氧性的临床应用

Clinical application of the oxygen transmissibility of powered contact lenses.

作者信息

Brennan N A, Efron N, Weissman B A, Harris M G

机构信息

Department of Optometry, University of Melbourne, Parkville, Victoria, Australia.

出版信息

CLAO J. 1991 Jul;17(3):169-72.

PMID:1893539
Abstract

Stipulation of the oxygen transmissibility (Dk/L) of a contact lens based upon oxygen permeability (Dk) and nominal lens thickness (L) can be misleading since center thickness, average thickness, and hence Dk/L of a contact lens vary with lens power. While earlier confusion regarding determination of Dk has now been resolved, this data has yet to be fully incorporated into a clinically applicable model. We developed a graphical solution describing variations in Dk/L with power for a representative sample of currently available hydrogel and rigid contact lenses. The Dk/L is found to be optimal for lenses of low minus power (approximately -2.50 D). The model presented here will assist practitioners in assessing the potential oxygen performance of different lens designs with respect to a given power. We advocate that lens manufacturers print the average Dk/L on lens packaging, according to the model described in this paper, along with the parameters that are usually stated.

摘要

基于透氧性(Dk)和标称镜片厚度(L)来规定隐形眼镜的氧透过率(Dk/L)可能会产生误导,因为隐形眼镜的中心厚度、平均厚度以及因此产生的Dk/L会随镜片度数而变化。虽然早期关于Dk测定的混淆现已得到解决,但该数据尚未完全纳入临床适用模型。我们开发了一种图形解决方案,描述了当前可用水凝胶和硬性隐形眼镜代表性样本的Dk/L随度数的变化情况。发现对于低负度数(约-2.50 D)的镜片,Dk/L最佳。本文提出的模型将帮助从业者评估不同镜片设计在给定度数下的潜在氧性能。我们主张镜片制造商根据本文所述模型,在镜片包装上打印平均Dk/L以及通常标注的参数。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验