植入角度对正畸微型种植体初期稳定性的影响

Insertion angle impact on primary stability of orthodontic mini-implants.

作者信息

Wilmes Benedict, Su Yu-Yu, Drescher Dieter

机构信息

Department of Orthodontics, University of Duesseldorf, Duesseldorf, Germany.

出版信息

Angle Orthod. 2008 Nov;78(6):1065-70. doi: 10.2319/100707-484.1.

Abstract

OBJECTIVE

To analyze the impact of the insertion angle on the primary stability of mini-implants.

MATERIALS AND METHODS

A total of 28 ilium bone segments of pigs were embedded in resin. Two different mini-implant sizes (Dual-Top Screw 1.6 x 8 mm and 2.0 x 10 mm) were inserted at seven different angles (30 degrees , 40 degrees , 50 degrees , 60 degrees , 70 degrees , 80 degrees , and 90 degrees ). The insertion torque was recorded to assess primary stability. In each bone, five Dual-Top Screws were used to compensate for differences in local bone quality.

RESULTS

The angle of mini-implant insertion had a significant impact on primary stability. The highest insertion torque values were measured at angles between 60 degrees and 70 degrees (63.8 degrees for Dual-Top 1.6 mm and 66.7 degrees for Dual-Top 2.0 mm). Very oblique insertion angles (30 degrees ) resulted in reduced primary stability.

CONCLUSIONS

To achieve the best primary stability, an insertion angle ranging from 60 degrees to 70 degrees is advisable. If the available space between two adjacent roots is small, a more oblique direction of insertion seems to be favorable to minimize the risk of root contact.

摘要

目的

分析植入角度对微型种植体初期稳定性的影响。

材料与方法

将28段猪髂骨块包埋于树脂中。两种不同尺寸的微型种植体(双尖螺钉1.6×8mm和2.0×10mm)以7种不同角度(30度、40度、50度、60度、70度、80度和90度)植入。记录植入扭矩以评估初期稳定性。在每块骨中使用5枚双尖螺钉以补偿局部骨质量的差异。

结果

微型种植体的植入角度对初期稳定性有显著影响。在60度至70度之间的角度测量到最高的植入扭矩值(双尖1.6mm为63.8度,双尖2.0mm为66.7度)。非常倾斜的植入角度(30度)会导致初期稳定性降低。

结论

为获得最佳的初期稳定性,建议植入角度在60度至70度之间。如果两个相邻牙根之间的可用空间较小,更倾斜的植入方向似乎有利于将牙根接触的风险降至最低。