Weber A, Rackelmann E
Landesuntersuchungsamt für das Gesundheitswesen Nordbayern, Nürnberg.
Berl Munch Tierarztl Wochenschr. 1991 Jul 1;104(7):233-5.
133 samples of food were investigated for comparison of a recent commercially available 1-2 test (Biocontrol) with a cultural standard method (non-selective pre-enrichment in buffered peptone water, selective enrichment in Rappaport-Vassiliadis-medium and selenite brilliant-green mannitol enrichment broth, inoculation on two selective agars) for presence of Salmonella. The 1-2 test showed in positive results an accuracy ("sensitivity") of 94.7% and in negative results an assurance ("specificity") of 97.9% and is therefore considered suitable for detection of Salmonella contaminated food.