松质骨标本压缩试验中杨氏模量的低估。

The underestimation of Young's modulus in compressive testing of cancellous bone specimens.

作者信息

Odgaard A, Linde F

机构信息

Biomechanics Laboratory, Orthopaedic Hospital, University of Aarhus, Denmark.

出版信息

J Biomech. 1991;24(8):691-8. doi: 10.1016/0021-9290(91)90333-i.

Abstract

In order to determine the accuracy of measurements of Young's modulus of cancellous bone by conventional compression testing, two independent strain measurements were made simultaneously during non-destructive uniaxial compression to 0.8% strain of rectangular specimens (n = 18). Strain was measured by an extensometer attached to the compression anvils close to the specimen and by an optical system covering the central half of the specimens. Mean Young's modulus determined by the extensometer technique was 689 MPa, but was 871 MPa when determined by the optical technique (mean difference = 182 MPa, SED = 50 MPa, p less than 0.002). Uneven strain distribution due to lack of support of cut vertical trabeculae at the anvil-specimen interface is believed to be causing the underestimation of Young's modulus measured by the extensometer technique. The influence of friction at the specimen-anvil interface was studied by performing a finite element analysis. It is concluded that Young's modulus of specimens of the chosen geometry on average is underestimated by about 20% by conventional compressing testing. The underestimation seems not to be dependent upon specimen density.

摘要

为了通过传统压缩试验确定松质骨杨氏模量测量的准确性,在对矩形标本(n = 18)进行无损单轴压缩至0.8%应变的过程中,同时进行了两次独立的应变测量。应变通过附着在靠近标本的压缩砧座上的引伸计以及覆盖标本中央一半的光学系统进行测量。引伸计技术测定的平均杨氏模量为689MPa,但光学技术测定时为871MPa(平均差异 = 182MPa,标准误 = 50MPa,p < 0.002)。据信,由于在砧座 - 标本界面处切割的垂直小梁缺乏支撑导致应变分布不均匀,从而造成引伸计技术测量的杨氏模量被低估。通过进行有限元分析研究了标本 - 砧座界面处摩擦的影响。得出的结论是,通过传统压缩试验,所选几何形状标本的杨氏模量平均被低估约20%。这种低估似乎与标本密度无关。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验