Department of Prosthetic Dentistry & Oral Implantology, Academic Centre for Dentistry (ACTA) Amsterdam, The Netherlands.
J Oral Rehabil. 2009 Apr;36(4):279-83. doi: 10.1111/j.1365-2842.2009.01944.x. Epub 2009 Feb 10.
The aim of this study was to determine the intra- and interobserver reliability and validity of the instrumental assessment of primary dental implant stability, using resonance frequency analysis (RFA). Sixteen tapered implants and 16 cylindrical implants were installed in eight unfixed dry human mandibles (Cawood classification IV/V). Implant stability quotients (ISQ; the outcome variable of RFA) and peak removal torque were determined. Both the intra-observer reliability and the interobserver reliability of the RFA measurements were fair-to-good, while no significant correlations between the ISQ values and removal torque were found. The removal torque of the cylindrical implants was higher than that of the tapered implants. The smallest detectable difference was almost nine ISQ units. Within the limitations of the present dry cadaver study, it was concluded that (i) primary dental implant stability can be assessed reliably with RFA measurements, (ii) the concurrent validity between RFA measurements and removal torque is poor, (iii) cylindrical implants may be more stable than tapered ones and (iv) two subsequent readings of RFA measurements need to differ at least nine ISQ units before the difference between the two measurements can be considered statistically significant. More research is needed to see whether these conclusions can be extrapolated to the clinical situation, including the assessment of implants during function (secondary stability).
本研究旨在使用共振频率分析(RFA)确定初步牙科种植体稳定性的仪器评估的内-间观察者可靠性和有效性。16 个锥形种植体和 16 个圆柱形种植体被植入 8 个未固定的干燥人类下颌骨(Cawood 分类 IV/V)中。测量了种植体稳定性比(RFA 的结果变量)和最大去除扭矩。RFA 测量的内-间观察者可靠性均为尚可-良好,而 ISQ 值与去除扭矩之间无显著相关性。圆柱形种植体的去除扭矩高于锥形种植体。最小可检测差异几乎为 9 个 ISQ 单位。在本干燥尸体研究的限制内,得出结论:(i)RFA 测量可可靠地评估初步牙科种植体稳定性,(ii)RFA 测量与去除扭矩之间的同时有效性较差,(iii)圆柱形种植体可能比锥形种植体更稳定,以及(iv)RFA 测量的两次连续读数需要相差至少 9 个 ISQ 单位,才能认为两次测量之间的差异具有统计学意义。需要进一步研究这些结论是否可以外推到临床情况,包括在功能(继发性稳定性)期间评估种植体。