Artifacts on the optic nerve head analysis of the optical coherence tomography in glaucomatous and nonglaucomatous eyes.

作者信息

Ortega Julio De León, Kakati Bobby, Girkin Christopher A

机构信息

Department of Ophthalmology, University of Alabama at Birmingham, USA.

出版信息

J Glaucoma. 2009 Mar;18(3):186-91. doi: 10.1097/IJG.0b013e31818159cb.

Abstract

PURPOSE

To determine the frequency and type of imaging artifacts in the optic nerve head (ONH) analysis protocol of the optical coherence tomograph (Stratus OCT) in healthy and glaucomatous eyes.

DESIGN

Retrospective study.

METHODS

Two experienced operators reviewed imaging obtained in 264 eyes from 264 participants (146 controls, 118 glaucoma patients). Each participant had a minimum of 3 scans using fast ONH protocol. Only scans with good quality were included in the study.

RESULTS

Thirty-two eyes (12%) had imaging artifacts in at least one of the radial scans distributed as follows: 12 eyes showed incorrect identification of vitreous-retinal surface plane, 3 eyes had incorrect positioning of retinal pigment epithelium-choriocapillaris border reference points, 10 eyes had vitreous tuft and/or remnants of posterior hyaloid artery above the optic disk that were mistakenly included in the estimation of optic disk parameters (rim and cup area), and 7 eyes with small optic disk were reported with rim area of 0 because of erroneous placement of reference planes.

CONCLUSIONS

In the Stratus OCT ONH protocol, careful inspection of each of the 6 radial scans is needed, as artifacts may be present in any of these scans. For the most common artifact found, modifications on retinal sensitivity may be successful in correcting such errors.

摘要

目的

确定在健康眼和青光眼眼中,光学相干断层扫描仪(Stratus OCT)的视神经乳头(ONH)分析协议中成像伪像的频率和类型。

设计

回顾性研究。

方法

两名经验丰富的操作人员回顾了264名参与者(146名对照者,118名青光眼患者)的264只眼睛的成像。每位参与者至少使用快速ONH协议进行了3次扫描。研究仅纳入质量良好的扫描。

结果

32只眼睛(12%)在至少一次径向扫描中存在成像伪像,分布如下:12只眼睛显示玻璃体-视网膜表面平面识别错误,3只眼睛视网膜色素上皮-脉络膜毛细血管边界参考点定位错误,10只眼睛视盘上方有玻璃体束和/或后玻璃体动脉残余,在视盘参数(边缘和杯状区域)估计中被错误纳入,7只小视盘眼睛因参考平面放置错误报告边缘区域为0。

结论

在Stratus OCT ONH协议中,需要仔细检查6次径向扫描中的每一次,因为这些扫描中的任何一次都可能存在伪像。对于发现的最常见伪像,调整视网膜敏感度可能成功纠正此类错误。