Suppr超能文献

利用光学相干断层扫描对高危人群进行青光眼筛查。

Screening for glaucoma in high-risk populations using optical coherence tomography.

机构信息

Department of Epidemiology, Biostatistics and Occupational Health, McGill University, Montreal, Quebec, Canada.

出版信息

Ophthalmology. 2010 Mar;117(3):453-61. doi: 10.1016/j.ophtha.2009.07.033. Epub 2010 Jan 19.

Abstract

OBJECTIVE

To estimate the diagnostic accuracy of Stratus optical coherence tomography (OCT) for glaucoma screening in high-risk populations.

DESIGN

Cross-sectional evaluation of a diagnostic test for screening.

PARTICIPANTS

Three hundred thirty-three community-based volunteer participants with risk factors for glaucoma.

METHODS

The optic nerve and peripapillary retinal nerve fiber layer (RNFL) of participants' eyes were scanned using the Stratus OCT. Based on an ophthalmologic examination and frequency doubling perimetry, eyes were classified into 4 categories: normal, possible glaucoma, probable glaucoma, and definitive glaucoma.

MAIN OUTCOME MEASURES

The sensitivities, specificities, positive and negative likelihood ratios of the RNFL, optic disc parameters, and their combinations were calculated.

RESULTS

The right eyes were retained for analyses. After excluding eyes with missing data or with poor quality scans, the data of 210 right eyes were analyzed. Six eyes had definitive glaucoma. Combining the best performing optic nerve head parameters (cup diameter or cup/disc vertical ratio or cup/disc area ratio) and RNFL parameters (superior average or inferior average or overall average) using AND-logic resulted in a sensitivity of 67% (95% confidence interval [CI], 24%-94%), specificity of 96% (95% CI, 92%-98%), a positive likelihood ratio of 17.08 (95% CI, 7.06-41.4), and a negative likelihood ratio of 0.35 (95% CI, 0.11-1.08).

CONCLUSIONS

When adequate quality scans may be obtained, the Stratus has moderate sensitivity and high specificity for definitive glaucoma. Specificity is increased when parameters from both the optic nerve head and RNFL scans are combined.

摘要

目的

评估 Stratus 光学相干断层扫描(OCT)在高危人群中进行青光眼筛查的诊断准确性。

设计

用于筛查的诊断测试的横断面评估。

参与者

333 名具有青光眼风险因素的社区志愿者参与者。

方法

使用 Stratus OCT 扫描参与者的视神经和视盘周围视网膜神经纤维层(RNFL)。根据眼科检查和频域加倍视野检查,将眼睛分为 4 类:正常、可能青光眼、可能青光眼和明确青光眼。

主要观察指标

计算 RNFL、视盘参数及其组合的敏感性、特异性、阳性和阴性似然比。

结果

保留右眼进行分析。排除数据缺失或扫描质量差的眼睛后,分析了 210 只右眼的数据。6 只眼患有明确的青光眼。使用 AND 逻辑结合表现最佳的视盘参数(杯直径或杯/盘垂直比或杯/盘面积比)和 RNFL 参数(上平均值或下平均值或总体平均值),敏感性为 67%(95%置信区间 [CI],24%-94%),特异性为 96%(95% CI,92%-98%),阳性似然比为 17.08(95% CI,7.06-41.4),阴性似然比为 0.35(95% CI,0.11-1.08)。

结论

当可以获得足够质量的扫描时,Stratus 对明确的青光眼具有中等敏感性和高特异性。当结合视神经头和 RNFL 扫描的参数时,特异性会提高。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验