[An evaluation of automated hematology analyzers in detecting abnormal blood cells].

作者信息

Tanaka Y, Miyachi H, Kawada T, Ono H, Gondo K, Seki T, Ikeda M, Ando Y

机构信息

Central Clinical Laboratory, Tokai University Hospital, Isehara.

出版信息

Rinsho Byori. 1991 Sep;39(9):961-6.

PMID:1942571
Abstract

We report the usefulness and capability of two hematology analyzers with different principles in characterization of abnormal cells such as immature granulocytes and blasts. We studied a total 1635 specimens which included 69 cases with hematopoietic malignancies. THMS H-1 (Technicon Instruments) and NE-8000 (Toa Instruments) were used as analyzers. An effort was made to determine the sensitivity and specificity of positive abnormal cell flag in comparison with the presence of abnormal cells seen on a manual differential. True positive rate for immature granulocytes in H-1 and NE were 43% and 30%, respectively. False positive rate for immature granulocytes in H-1 and NE were 0.01% and 0.03%, respectively. With both analysers, cases with low percentage of immature granulocytes generated negative flag for abnormal cells. True positive rate for blasts in H-1 and NE were 74% and 65%, respectively. False positive rate for blasts in H-1 and NE were 1.8% and 0.2%, respectively. With both instruments, in all cases with blast count over 100/microliters, blasts were detected. Only H-1 detected the cases with blast counts less than 100/microliters. In H-1 all cases with ATL, one case with AMMoL and another with ALL generated negative blast flag. Upon morphologic examination, these blasts were found to be large and peroxidase negative with convoluted and irregularly shaped nuclei. On the other hand, NE detected blasts regardless of those morphology. In conclusion, Both instruments are useful in screening abnormal blood cells, while their unique capability must be considered.

摘要

相似文献

5
Blast flagging with the UniCel DxH 800 Coulter Cellular Analysis System.
Lab Hematol. 2010 Jun;16(2):23-5. doi: 10.1532/LH96.09015.
6
Differences in detecting blasts between ADVIA 2120 and Beckman-Coulter LH750 hematology analyzers.
Int J Lab Hematol. 2010 Feb;32(1 Pt 2):113-6. doi: 10.1111/j.1751-553X.2008.01113.x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验