Bon Pierre, Maucort Guillaume, Wattellier Benoit, Monneret Serge
Aix-Marseille Université, Ecole Centrale Marseille, CNRS, Institut Fresnel, Campus de Saint-Jérôme, 13013 Marseille, France.
Opt Express. 2009 Jul 20;17(15):13080-94. doi: 10.1364/oe.17.013080.
Phase imaging with a high-resolution wavefront sensor is considered. This is based on a quadriwave lateral shearing interferometer mounted on a non-modified transmission white-light microscope. The measurement technology is explained both in the scope of wave optics and geometrical optics in order to discuss its implementation on a conventional microscope. In particular we consider the effect of a non spatially coherent source on the phase-image signal-to-noise ratio. Precise measurements of the phase-shift introduced by microscopic beads or giant unilamellar vesicles validate the principle and show the accuracy of the methods. Diffraction limited images of living COS-7 cells are then presented, with a particular focus on the membrane and organelle dynamics.
考虑使用高分辨率波前传感器进行相位成像。这基于安装在未改装的透射式白光显微镜上的四波横向剪切干涉仪。为了讨论其在传统显微镜上的实现方式,在波动光学和几何光学的范围内对测量技术进行了解释。特别地,我们考虑了非空间相干源对相图像信噪比的影响。由微珠或巨型单层囊泡引入的相移的精确测量验证了该原理,并展示了这些方法的准确性。随后展示了活COS - 7细胞的衍射极限图像,特别关注了膜和细胞器的动态变化。