Analytical Chemistry Department, Faculty of Sciences, University of Burgos, Burgos, Spain.
Biosens Bioelectron. 2010 Feb 15;25(6):1333-7. doi: 10.1016/j.bios.2009.10.024. Epub 2009 Oct 27.
Horseradish peroxidase has been successfully immobilized in a polypyrrole matrix onto disposable screen-printed carbon electrodes for the selective detection of Ochratoxin A. The chronoamperometric determination of this mycotoxin has been optimized by experimental design methodology, which implied the join evaluation of pH of the buffer solution, applied potential and concentration of H(2)O(2). The slopes of the calibration curves built under the optimum conditions of the experimental variables have been used to estimate the reproducibility and the repeatability of the developed biosensor for Ochratoxin A. Relative standard deviation values of 1.9% (n=5 and alpha=0.05) and 7.1% (n=4 and alpha=0.05) were obtained for reproducibility and repeatability, respectively. The capability of detection for this method was 0.1 ng mL(-1) (alpha=0.05 and beta<0.05). The viability of the developed biosensor in the determination of Ochratoxin A in spiked beer and in roasted coffee has been shown, yielding average recoveries of 103% and 99%, in that order, with relative standard deviation values less than 5% (n=3 and alpha=0.05) in both cases.
辣根过氧化物酶已成功固定在聚吡咯基质上,用于在一次性丝网印刷碳电极上选择性检测赭曲霉毒素 A。通过实验设计方法优化了该真菌毒素的计时安培法测定,该方法涉及缓冲溶液 pH 值、施加电位和 H2O2 浓度的联合评估。在实验变量最佳条件下构建的校准曲线的斜率用于估计开发的赭曲霉毒素 A 生物传感器的重现性和重复性。重现性和重复性的相对标准偏差值分别为 1.9%(n=5,alpha=0.05)和 7.1%(n=4,alpha=0.05)。该方法的检测能力为 0.1 ng mL(-1)(alpha=0.05 和 beta<0.05)。已经证明了开发的生物传感器在检测加标啤酒和烤咖啡中的赭曲霉毒素 A 的可行性,在这两种情况下,平均回收率分别为 103%和 99%,相对标准偏差值均小于 5%(n=3,alpha=0.05)。